36
Use the following troubleshooting guide to help solve problems with your NeuRx
TM
EPG:
Problem
Action
Pacing of the diaphragm stops.
Check the connection of the patient
cable to the electrode connector.
Check the connection of the patient
cable to the NeuRx
®
EPG.
Try a different patient cable.
Inspect the connections of the
electrode wires to the electrode
connector for breakage.
Discomfort while pacing.
STOP
use of the NeuRx
®
EPG and call
your healthcare provider who is helping
you with your NeuRx DPS
®
. The EPG
settings may need adjustment.
Bleeding, bruising, or infection where the
electrode wires pass through the skin.
Call your healthcare provider who is
helping you with your NeuRx DPS
®
.
Pain at the electrode wire exit site.
STOP
use of the NeuRx
®
EPG and call
your healthcare provider who is helping
you with your NeuRx DPS
®
.
Skin irritation or hypersensitivity to
stimulation.
Call your healthcare provider who is
helping you with your NeuRx DPS
®
.
Multiple “
X
”s appear on the NeuRx
®
EPG
screen.
Try a different patient cable. If the
problem continues, disconnect the
patient cable from the NeuRx
®
EPG and
insert the test plug (found in the patient
kit). If the EPG is working properly, use
the Backup Indifferent Electrode
Interconnect. If the problem continues,
call your healthcare provider who is
helping you with your NeuRx DPS
®
.
The NeuRx
®
EPG is exposed to
substantial amount of water or fluid.
STOP
use of the NeuRx
®
EPG. Call your
healthcare provider who is helping you
with your NeuRx DPS
®
.
Alarm sounds during each stimulation.
Check the connection of the patient
cable to the electrode connector.
Check the connection of the patient
cable to the NeuRx
®
EPG.
Try a different patient cable.
Inspect the connections of the
electrode wires to the electrode
connector for breakage.
The NeuRx
®
EPG beeps for 10 seconds
every hour. (This means the NeuRx
®
EPG is running on its backup battery.)
Replace the main battery as described in
this manual. See the Battery Installation
section of this manual for the steps.
Содержание NeuRx
Страница 27: ...27 Figure 13 Replacing the disposable battery ...
Страница 46: ...46 ...
Страница 47: ...47 ...