The verification can now be made with the program
SYCOPRO for WINDOWS
FROM SYLVAC,
which constructs a complete plotting in both directions.
7.3 STANDARD PROCEDURE
Once the device has been adjusted, and if you wish to verify a different indicator,
proceed as follows:
Undo the indicator locking nut.
Rotate the mechanism 90°, in order to extract the reference ball from the measuring
fork.
Remove the indicator
Inspect the position of the support, and fix it at the same position on the dove’s tail of
the next indicator to be verified.
Engage the support in the mechanism.
Rotate the mechanism to bring the reference ball to the centre of the measuring fork
Tighten the locking nut.