22
Standard delivery
In the event of problems
Inactive corrections indicator :
•
Reactivate the corrections as follows
1.
RS232 COR ON command
2.
Retake the Ref
Switch the instrument on by pressing
ON + SETUP
Wait for the «init» display then browse
using the keys after instrument
initialisation.
Important: This action must be done only with the base key (3mm ball)
Settings menu :
Check the moving carriage alignment sensor :
Taking the reference
Lin correction, in µm/300mm
Test the UP/DOWN button; active 100% 4.0 - 4.1 V
Option to modify the motor release step to measure small
diameters with the Left/Right buttons.
Rest :
2.2 - 2.4 V
UP :
2.8 - 3.0 V
DOWN :
1.6 - 1.8
V
Hi_Cal 150 mm
Réf. N° : 830.0150
Hi_Cal 300 mm
Réf N° : 830.0300
Hi_Cal 450 mm
Réf N° : 830.0450
Including:
•
1 fixed probe, with saphire ball 3 mm
•
1 master gauge
•
1 charger 9 V | 1100mA (mains AC voltage according to the country)
•
1 protection cover
•
1 calibration certificat
•
1 instruction manuel
•
Other accessories: see the site www.sylvac.ch