Operating Instructions
22
2.
Calibration
(depending on the selection in system 7.2 Sensors)
2.1
Calibration factor
input:
0.01 … 9.99
Factor for the subsequent adjustment of the
actual measurement. All measurements
are measured with this factor.
2.2
Calibration filter
Input:
n = 1 … 9999
Average number n for the raw value when
performing a calibration. n values are
combined into an average value with PT1
character, in order to obtain a more stable
representation.
This filter also affects the representation of
the calculated raw value in the online
window.
2.3
Calibration points
(support points)
Input:
2 … 5
Number of support points for a
linearisation above the operating range.
2.4
Calibration
Calibration sub-menu
2.4.1 P1 value
Input:
Measurement
Output measurement in the selected
mass/time unit.
(2.4.2) P1 calibration
Transfer:
Raw value
Transfer of the current raw value (filtered)
from the mass flow with the key [
←
].
The value can also be entered directly.
. . . (depending on the number of support points)
For additional support points (depending on
[ 2.3 ]), additional value pairs can be set.
2.4.n Pn value
Input:
Measurement to be displayed
(2.4.n) Pn calibration
Transfer:
Current raw value