MODEL 2200-3000  RETROFIT OPEN STREAM CURRENT METER / PAGE 3 of 7 

Supplement to Model 3000 Instruction Manual

 

 

7.  Conduct the standard AA meter spin test as defined by the USGS instructions.  When properly 

installed and adjusted, the AA meter conversion should provide at least a slight increase in the normal 
spinning time. 

8.  Install the "O" Ring (2200-011) in the top of the 2200-A2 Adapter. 
9.  Place the Sensor (2200-A) on the adapter and tighten the sensor mounting nut (2200-A3) finger tight 

only.  The sensor should not rotate when mounting nut is tightened.  If it does tighten the front yoke 
screw a bit more (see step 4 above) and check that the "O" ring is properly seated. 

10. Give the meter a second spin test to make sure all installations are correct and that operation is 

satisfactory. 

PYGMY METER 

1.  Remove contact chamber cap, binding post nut and binding post body, leaving only the nylon bushing 

in place in the yoke.  Install the furnished 

10-32 x 1/4"

 screw tightly in the bushing. 

2.  Remove the bucketwheel pivot shaft. 
3.  Unscrew the upper rotor shaft from the bucketwheel and replace it with the new one 

(2200-A5)

.  

Tighten the new shaft only by using the wrench 

(.05" allen wrench)

 inserted in the hole in the rotor 

shaft to avoid damage to the bearing surface. 

4.  Remove Upper Bearing (bronze part) from the yoke.  This part is not necessary and will prevent 

proper installation of the adapter if not removed. 

5.  Install the Pygmy meter Adapter 

(2200-A1)

 by threading it into the upper yoke arm (formerly the 

contact chamber cap threads).  Carefully insert the top of the new bucketwheel shaft through the hole 
in the adapter bottom while you thread.  Tighten the adapter onto the yoke by use of the allen wrench 
horizontally through the holes in the upper threads of the adapter.  Make sure the adapter is tight to the 
yoke otherwise removal of the sensor nut may unscrew the adapter if the threads are a bit snug. 

6.  Re-install the lower pivot shaft adjusting it vertically as necessary; centering the bucketwheel in the 

yoke as much as possible. 

7.  Using the brass blade driver supplied install the Fiber-Optic Carrier Rotor 

(2200-A14)

 through the top 

of the Adapter and onto the threads of the new rotor shaft.  DO NOT TIGHTEN TOO TIGHT TO 
AVOID STRIPPING THREADS.  

The top of the Fiber-Optic Rotor must not be above the 

shoulder inside the Adapter.

 

8.  Conduct a spin test.  When properly installed and adjusted, the spinning time should be at least as long 

as that previously achieved before conversion. Note the rotation of the fiber-optic rotor. It should be in 
the center of the adapter and must never come in contact with the sides of the adapter. 

See Pygmy 

Meter Retrofit Problems, ahead, and refer to the notes for corrections if contact is being made. Do not 
operate the meter if fiber-optics rotor makes contact with the adapter!

 

9.  Install the "O" Ring 

(2200-011)

 in the top of the Adapter and place the Sensor on the Adapter and 

tighten the Sensor Retaining Nut 

(2200-A3)

 on the yoke finger tight only. Do not operate the sensor 

without an “O” ring as it helps keep the optics from contacting the rotor should tolerances be wrong. 

10. Conduct a second spin test to confirm all installations and adjustments are satisfactory. 

 

NOTE:

 

Keep the adapter protectors 

(2200-A16)

 on the meters when the instrument is not in 

use.  These protect the adapter threads, prevent loss of the "O" rings and keep dirt out of the 
sensor.

 

CARE AND MAINTENANCE OF THE MODEL 2200-3000 SENSOR 

1.  The proper operation of the Model 2200 Sensor depends on the transmission of an infrared light beam 

generated by the light-emitting diode in the Sensor through 180° loops of rotating fiber-optic bundles 

Содержание 2200-3000

Страница 1: ...red by the USGS Price Type Meters have been mechanically reliable for many years however accuracy has been a function of careful maintenance and adjustment of the delicate cats whisker contactors and...

Страница 2: ...ALLATIONS refer to drawings in back AA METER 1 Remove contact chamber assembly entirely 2 Remove pivot shaft from yoke at lower end of the bucket wheel 3 Unscrew and remove upper rotor shaft from the...

Страница 3: ...oles in the upper threads of the adapter Make sure the adapter is tight to the yoke otherwise removal of the sensor nut may unscrew the adapter if the threads are a bit snug 6 Re install the lower piv...

Страница 4: ...he Model 3000 Indicator operating instructions for further information on calibration and calibration corrections AA METER RETROFIT PROBLEMS 1 In some AA Yokes it has been found that a deep chamfer ha...

Страница 5: ...Indicator in the Count mode and spin the meter to be sure that the sensor is producing pulses With the normal two hole fiber optic rotor you should get two pulses per meter revolution If confirmed th...

Страница 6: ...7 Supplement to Model 3000 Instruction Manual 2200 A Sensor A3 RETAINING NUT 011 O RING A14 FIBER OPTICS CARRIER ASSEMBLY A2 YOKE ADAPTER 250 10 32 R H M SCREW A4 SHAFT YOKE 3 32 05 allen wrench use...

Страница 7: ...nt to Model 3000 Instruction Manual 2200 A Sensor Cable assembly A3 RETAINING NUT 011 O RING A14 FIBER OPTICS CARRIER ASSEMBLY A1 YOKE ADAPTER PYGMY A5 SHAFT 10 32 x 1 4 screw 3 32 05 Allen Wrench Use...

Отзывы: