MEASUREMENT 

1. 

Select in Swema 3000md, the external pressure sensor. The letters 

“INT.” are shown on the 

left at the lower part of the display. It is also possible to measure with the internal pressure 
sensor, but then the pressure hoses (silicone hoses) must be changed, and the letters 

“EXT.” 

are shown instead. 

2. 

Before measuring starts, both differential pressure sensors must be zeroed. Press 

“INT.” to 

access the built-in pressure sensor in Swema 3000md, and press ZERO. Afterwards, press 
“EXT.” to return to the externally connected pressure sensor and ZERO it. 

3. 

In measuring mode the display shows 4 measuring parameters. The largest digits at the top 
show the leakage flow (l/s, m

3

/h, l/sm

2

 or m

3

/hm

2

), below the leakage flow, the pressure drop 

over the orifice plate in the flange used to calculate the leakage flow. The two pressures 
shown below the pressure drop, and with small digits, are the expected pressure value at the 
left and the read value (actual value) of the measured pressure to the right (See picture 3). 

4. 

Press the arrow keys UP/DOWN to adjust the value of the pressure to be measured, wait until 
Swema 3000md is stabilized at the read value (true value) of the pressure. (See picture 3). 

5. 

Measure by pressing the ENTER button.

 

 

 

 

 