42
A-96.730.121 / 160920
Chematest 35
Menu Explanations
3.4 Set Time
Adjust date and time.
4 Settings
4.1 Sensors
Photometer
Enter the reference values of the verification cuvettes according to
the enclosed calibration certificate.
Range: 0.000–2.000
Electrode
pH
Filter time
constant
Used to damp noisy signals. The higher the filter time constant, the
slower the system reacts to changes of the measured value.
Range: 5–300 s
Temp. Comp.
Choose the compensation model which fits best to your application.
None
: no temperature compensation.
Nernst
: for potable water, waste water, swimming pools.
Coefficient
: for special applications.
Range: −0.100–0.100 pH unit per °C
NOTICE:
For this setting the unit °C is always used, regardless
of the setting under
Standards
A temperature curve is programmed for SWAN standard 1, pH 7
and SWAN standard 2, pH 9. If you want to use your own standards
you can readjust the temperature curve according to your stan
-
dards.
Standard 1
: Assign the measured pH value to the according
temperature from 0–50 °C in steps of 5 °C.
Standard 2
: Assign the measured pH value to the according
temperature from 0–50 °C in steps of 5 °C.
Electrode
Redox/ORP
Filter time
constant
Used to damp noisy signals. The higher the filter time constant, the
slower the system reacts to changes of the measured value.
Range: 5–300 s
Temp. Comp.
None
Nernst
Coefficient