Chematest 35
Version 2.02 and higher
A-96.730.121 / 160920
Operator’s Manual
Страница 1: ...Chematest 35 Version 2 02 and higher A 96 730 121 160920 Operator s Manual...
Страница 2: ...around the world For any technical question contact your nearest SWAN representative or the manufacturer SWAN ANALYTISCHE INSTRUMENTE AG Studbachstrasse 13 8340 Hinwil Switzerland Internet www swan c...
Страница 3: ...Rules for Photometric Measurements 12 4 2 How To Use Reagents 13 4 3 Notes on Individual Methods 14 4 4 Performing a Photometric Measurement 15 4 5 Automatic Shutoff 18 5 pH and Redox 19 5 1 Basic Rul...
Страница 4: ...tivity Sensor 33 9 7 Setting Time and Date 33 10 Troubleshooting 34 10 1 Instrument Errors 34 10 2 Photometric Measurements 35 10 3 Sensor Based Measurements 37 11 Menu Explanations 39 1 Data Storage...
Страница 5: ...Warning notices The symbols used for safety related notices have the following meaning WARNING Severe injuries or damage to the equipment can occur if such warnings are ignored Follow the prevention...
Страница 6: ...e instrument using a standard type A USB socket and the supplied USB cable Protect from heat and splash water during charging When no charging or sensor cable is connected close both sockets with the...
Страница 7: ...test 35 Product Description A 96 730 121 160920 5 2 Product Description 2 1 Instrument Overview A B C D Sensor socket Charging socket Cuvette adapter Display E F G Keypad Cuvette cap Cuvette A B C D E...
Страница 8: ...or off the instrument B ARROW UP to move up in a menu list and to increase digits C ARROW DOWN to move down in a menu list and to decrease digits D EXIT to exit a menu or command to move back to the p...
Страница 9: ...elected from the method catalogue B default setting or three methods that are defined by the user To define the methods that appear in the favorites list proceed according to Initial Setup p 9 B Metho...
Страница 10: ...t est 35 and a mobile device or PC The app includes the following functions Transfer of the stored measurements and export to a CSV or Excel file Editing the user list Editing the ID list Switching be...
Страница 11: ...enu Maintenance Set Time 3 3 Setting up ID and User List Selecting the attributes to be saved Navigate to Menu Settings Identification Attribute and select which attributes should be saved with each m...
Страница 12: ...ured to display the last three methods selected from the method catalogue B in the favorites list With this configuration the favorites list is automatically adjust ed each time a method is selected f...
Страница 13: ...subsection 4 Settings p 42 3 7 Sensor Calibration Before first use calibrate your pH and redox sensors See Calibration of the pH Sensor p 29 and Calibration of the Redox Sensor p 30 NOTICE The rubber...
Страница 14: ...ent wipe it off and carry out the measurement as quickly as possible Observe proper sampling Take the water samples about 10 cm below the water surface and about 50 cm from the edge of the pool Allow...
Страница 15: ...he drop counter and shake firmly until the DPD powder has completely dissolved Write the mixing date on the bottle The prepared reagent can be stored at room temperature for 2 months The reagents OXYC...
Страница 16: ...suring combined chlorine with one cuvette or with two cuvettes Apart from the fact that one or two cuvettes are used the two meth ods are identical Measuring with only one cuvette is faster and re qui...
Страница 17: ...surement with the Chematest 35 consists of the following steps Step 1 Zero measurement Step 2 Addition of reagents and subsequent measurement for some methods with further intermediate steps and or re...
Страница 18: ...to continue 83 Free chlorine Place the adapter in the Chematest and close the cap Enter for zero 83 Free chlorine Measuring zero Wait for instructions 83 Free chlorine Remove adapter with cuvette and...
Страница 19: ...E If you press Exit at this point you return to the main screen saving the measurement without user and ID Select Enter to continue and confirm with Enter The ID and the user are added to the entry in...
Страница 20: ...ser input e g when the mea surement result is displayed and no key is pressed for 10 minutes the instrument shuts off automatically to save power This affects the storage of measurement results in the...
Страница 21: ...an water To prevent the sensor from drying out fill the protection cap with clean water Then slide the knurled nut C onto the sensor shaft push the sen sor shaft into the protection cap as far as it w...
Страница 22: ...ently several times before letting it stand 6 Wait until both the pH and the temperature value are stable Then press Enter to save the result 7 Next you can optionally select an ID and or a user name...
Страница 23: ...abilized switch it on again and select the Redox ORP method again This will not affect the stabilization of the sensor 7 Next you can optionally select an ID and or a user name This procedure is descr...
Страница 24: ...d the sample water must have the same temperature The process of temperature equalization can take up to five minutes depending on the tem perature difference To accelerate this process it is helpful...
Страница 25: ...sensor stand in the sample Make sure that the vent holes A are a few millimeters below the water surface 6 Select Conductivity from the main screen or from the method catalogue The displayed conductiv...
Страница 26: ...user input e g when the mea surement result is displayed and no key is pressed for 10 minutes the instrument shuts off automatically to save power This affects the storage of measurement results in th...
Страница 27: ...5 Add Oxycon CH drop by drop and count the exact number of drops Mix after each drop and check the sample color First the sample color changes to light gray After the required number of drops has been...
Страница 28: ...can be transferred to a mobile device or PC and converted into an Excel or CSV file using the CT App Press the ARROW DOWN and ARROW UP keys to scroll through the data history 4 83 Data History Date T...
Страница 29: ...rence values for the cuvettes Low and High in Menu Settings Sensors Photometer Verification procedure Before each use check the expiration date on the calibration certif icate If the expiration date h...
Страница 30: ...the unit to the rear loosen the two screws A and remove the cover B Use a soft lint free cloth moistened with water to clean the photometer compartment Outside of the housing If reagents drip onto th...
Страница 31: ...Procedure 1 Plug the sensor cable into the sensor socket 2 Select Menu Maintenance Calibration Electrode pH and follow the instructions on the screen Calibration errors If the calibration is aborted w...
Страница 32: ...edox ORP and follow the instructions on the screen Calibration errors If the calibration is aborted with the message Offset error Make sure that the correct standard solution has been measured Careful...
Страница 33: ...ibration error Make sure that the correct standard solution has been measured Clean the sensor as described in Cleaning the Conductivity Sensor p 33 Repeat the calibration with a new standard solution...
Страница 34: ...eparations Make the following settings before performing the verification mea surement 1 From Menu Settings Sensors Conductivity Temp comp select the setting Coefficient 2 Set the coefficient to 1 95...
Страница 35: ...household detergent 3 Rinse the sensing tip and the protection sleeve with clean water 4 Screw the protection sleeve finger tight onto the sensor 9 7 Setting Time and Date The correct setting of the c...
Страница 36: ...rms Pending errors to see the error message Error Description Corrective action E001 Bluetooth Check if the installed firmware version supports data exchange with the CT App See Appendix CT App p 55 f...
Страница 37: ...ction Error E002 E003 or E004 is present See Instrument Errors p 34 for details Possible cause Corrective action Extraneous light Make sure that the cuvette cap is properly closed Cuvette dirty Clean...
Страница 38: ...ive action Extraneous light Make sure that the cuvette cap is properly closed Cuvette dirty Clean and rinse the cuvette Concentration or pH value outside the measuring range of the instrument Consult...
Страница 39: ...uctivity sensor Possible cause Corrective action Error E002 E003 or E004 is present See Instrument Errors p 34 for details Possible cause Corrective action Sensor not connected Check if the sensor is...
Страница 40: ...ible Sensor failure Consult the manual is displayed Possible cause Corrective action Cable or sensor defective Call support The sensor was disconnected from the instrument too early Do not unplug the...
Страница 41: ...ata Date time Measured value s ID User If the fields ID and User contain the text Empty the assignment of the attributes to the measurement was skipped by the user or de activated under 4 2 Identifica...
Страница 42: ...calibrations are memorized for each sensor Then the old est verification is deleted to save the newest one 2 Diagnostics 2 1 Alarms Pending errors Provides the list of active errors with their status...
Страница 43: ...identification feature during Bluetooth communication Firmware Firmware version of the Bluetooth module Hardware Hardware version of the Bluetooth module 3 Maintenance 3 1 Simulation For test purpose...
Страница 44: ...ls Coefficient for special applications Range 0 100 0 100 pH unit per C NOTICE For this setting the unit C is always used regardless of the setting under 4 3 Units p 44 Standards A temperature curve i...
Страница 45: ...d to damp noisy signals The higher the filter time constant the slower the system reacts to changes of the measured value Range 5 300 s Factor TDS Factor for the calculation of TDS Range 0 000 10 000...
Страница 46: ...hoice of languages varies depending on the installed language pack Requires hardware V2 0 or higher If the required language is not available on the Chematest 35 the CT App can be used to install a di...
Страница 47: ...mode Choose between detailed instructions and expert mode for fast workflow Method save Defines the behavior of the favorites list on the main screen For details see Initial Setup p 9 No The favorite...
Страница 48: ...46 A 96 730 121 160920 Chematest 35 Menu Explanations 4 5 Bluetooth Bluetooth Activates the Bluetooth connection to connect to the CT App Bluetooth Active Inactive...
Страница 49: ...al chlorine Combined chlorine A 85 590 400 Oxycon Des Free chlorine Free chlorine in presence of chlorine dioxide or bromine Chlorine dioxide Bromine Iodine A 85 590 500 Oxycon Ozone Ozone Ozone in pr...
Страница 50: ...475 mV 40 ml A 85 131 700 Oxycon RC 1 413 calibration solution for conductivity sensor 1 l Part no Description A 85 153 580 Veri Kit Transmission A 87 160 014 Swansensor pH CT A 87 460 014 Swansensor...
Страница 51: ...Numbers A 96 730 121 160920 49 C 85 520 040 Cleaning wipes for cuvette 100 pack A 70 065 600 Brush for cuvette cleaning C 88 917 200 USB charging cable C 89 511 010 Blister pack for reagent bottles Pa...
Страница 52: ...67 Charging time approximately 3 hours Battery life approximately 5000 measurements or one week of standby operation Automatic shutoff after 10 minutes of inactivity Conditions for operation Temperatu...
Страница 53: ...tc combined chlorine cc Combined chlorine using two cuvettes 0 10 ppm Free chlorine fac total chlorine tc combined chlorine cc Free chlorine in presence of chlorine dioxide or bromine 0 10 ppm Free c...
Страница 54: ...dox CT Pt redox electrode with integrated temperature sensor for consis tent data records Measuring range Resolution 400 1200 mV 1 mV Common characteristics Poisoning protected reference systems with...
Страница 55: ...ensation Connection 0 00 9 99 S cm 10 0 99 9 S cm 100 999 S cm 1 00 9 99 mS cm 10 0 29 9 mS cm 30 100 mS cm 0 2 S cm 1 5 of reading 0 50 C absolute none linear coefficient in C non linear function for...
Страница 56: ...rtment containing USB charging cable and 6 wipes for cuvette cleaning 2 cuvette adapters Dropper bottle for rinsing water F G H I J K 2 sensor compartments sensors sold separately 1 sensor protection...
Страница 57: ...p by clicking the appropriate icon or menu item in the browser On a Windows PC click the icon in the title bar Confirm by clicking Install in the next dialog On an Android device click Add Swan CT App...
Страница 58: ...modify the user list of the Chematest 35 C Edit locations Allows to modify the ID list of the Chematest 35 D Settings Here you can install a different language pack select a language or disconnect fro...
Страница 59: ...p under Change language For this the Chematest 35 must be connected to the CT App Select the desired language from the list If the language is part of the currently installed language pack it can be s...
Страница 60: ...luetooth module which has been tested and found to comply with the limits for a Class B digital device pursuant to part 15 of the FCC Rules Contains FCC ID T9JRN4020 This product meets all requirement...
Страница 61: ...Chematest 35 A 96 730 121 160920 59...
Страница 62: ...companies and distributors cooperates with independent representatives all over the world SWAN Products Analytical Instruments for High Purity Water Feedwater Steam and Condensate Potable Water Pool...