22
A-96.730.101 / 151118
Chematest 30
Troubleshooting
8.
Troubleshooting
This chapter contains some hints to facilitate troubleshooting. For
detailed information on handling and cleaning parts, see
. For detailed information on programming the instru
-
ment, see
8.1. Error List
Error
Description
Corrective action
E001
Bluetooth
– Switch the instrument on and off and check
if the error disappears.
– If the error persists, call support.
E002
IC ADC
– Call support.
E003
Factory data
– The instrument has lost its calibration data.
Send the instrument back to Swan.
E004
Invalid time
– Set time and date under <Menu>/
<Maintenance>/<Set Time>.
E017
Events erased
– No action necessary.
– This message informs that the measured
data have been deleted by calling the
function "Set to factory defaults”.