Installation
30
3
Fix the KCl bottle upside down in the holder on the panel.
4
Puncture the bottle bottom to allow pressure equilibration.
5
Knock against the KCl bottle to remove trapped air bubbles [B] in
the dosing tip.
Note:
Air bubbles trapped in the dosing tip of the KCl bottle may
stop the KCl flow to the reference electrode, which results in
wrong measuring values.
6
Slip the union screw [J], and washer [K] on the electrode shaft [I].
7
Wet the O-ring [L] and slip it carefully over the electrode shaft [I].
8
Insert the electrode through the hole [H] into the reference cham
-
ber and push it down until the ground joint diaphragm is about 0.5
cm above the bottom.
9
Tighten the union screw [J] finger tight.
10
Remove the connector cap [F] from the electrode.
11
Screw the connector [E], of the cable marked with R, onto the
electrode.
12
Connect the cable to the front end in the AMI Transmitter, see
.
Extension tube
on compact
panel version
On the compact panel version, use the enclosed extension tube [C]
and the connecting piece [D] to connect the KCl bottle.
A
B
C
D
E
KCl bottle
Dosing tip
Extension tube
Connecting piece
KCl supply tube
A
B
C
D
E