A-96.250.211 / 171121
AMI Sodium P
Installation
27
3.4.2
Installing the Reference Electrode
General
The SWAN reference electrode is a double junction Calomel / KCl
type electrode. The outer liquid junction is a liquid glass sleeve, guar
-
anteeing easy maintenance and long life time.
To maintain correct operation make sure that:
the ground joint diaphragm stays clean and a KCl flow of about
1ml/day is maintained.
no air bubbles are trapped in the electrode and in the tube to
the KCl reservoir.
the electrical connectors stay clean and dry.
CAUTION
KCl is corrosive
Do not let electrolyte drop on the measuring cell.
Prepare the
KCl Bottle
1
Remove the seal cap [A] from the dosing tip [B].
2
Cut off the upper sealed part of the dosing tip.
A
B
C
Seal cap
Dosing tip
KCl bottle
A
B
C