AMI Deltacon Power
Program List and Explanations
74
5.3.2.32.31
Control Parameters
Range for each Parameter same as
Dosing is controlled by the position of a motor driven mixing valve.
5.3.2.32.22
Run time:
Time needed to open a completely closed valve
Range: 5–300 Sec.
5.3.2.32.32
Neutral zone:
Minimal response time in % of the runtime. If the re
-
quested dosing output is smaller than the response time, no change
will take place.
Range: 1–20 %
5.3.2.32.4
Control Parameters
Range for each Parameter same as
The relay will be activated repetitively depending on the programmed
time scheme.
5.3.2.24
Mode:
Operating mode (interval, daily, weekly)
5.3.2.340
Interval/Start time/Calendar: Dependent on options operating mode.
5.3.2.44
Run time:
time the relay stays active.
Range: 5–32’400 Sec
5.3.2.54
Delay:
during run time plus the delay time the signal and control out
-
puts are held in the operating mode programmed below.
Range: 0–6’000 Sec
5.3.2.6
Signal Outputs:
select the behavior of the signal outputs when the re
-
lay closes. Available values: cont., hold, off
5.3.2.7
Output/Control:
select the behavior of the control outputs when the
relay closes. Available values: cont., hold, off
The relay will be switched via the Profibus input. No further parame
-
ters are needed.
5.3.4
Input:
The functions of the relays and signal outputs can be defined
depending on the position of the input contact, i.e. no function,
closed or open.
5.3.2.32.1
Actuator = Motor valve
5.3.2.1
Function = Timer:
5.3.2.1
Function = Fieldbus: