A-96.250.451 / 190121
AMI Deltacon Power
Program List and Explanations
62
4.4.2
Screen 2
Same as screen 1.
5 Installation
5.1 Sensors
5.1.1
Miscellaneous
:
5.1.1.1
Calculations:
Select “yes” if pH and ammonia concentrations should
be calculated. pH and ammonia are now available on screen 1 or 2,
on the signal outputs and as alarm or limit values.
5.1.1.2
Meas. unit:
Choose the measuring unit as µS/cm or µS/m
5.1.1.3
Monitoring of resin:
Select “yes” if consumption of cation resin should
be calculated and displayed. Replacement of exhausted resin must
now be confirmed in "Maintenance".
5.1.1.4
Resin Capacity:
Enter the resin capacity.
Range: 0.5–4.0 eq/l
5.1.1.5
Volume of resin:
Enter the volume of the resin bottle.
Range: 0.5–30.0 l
5.1.2
Sensor parameters
:
5.1.2.1
Sensor 1
5.1.2.1.1
Cell Constant
: Enter the cell constant printed on the sensor label.
5.1.2.1.2
Temp. Corr
: Enter the temperature correction printed on the sensor
label.
5.1.2.1.3
Cable length
: Enter the cable length. Set the cable length to 0.0 m if
the sensors are installed in the flow cell on the AMI monitor.
5.1.2.1.5
Temp. comp
:
5.1.2.1.5.1
Comp
.: Available compensation models:
Strong acids (Never select strong acids for sensor 1!)
Strong bases
Ammonia
Morpholine
Ethanolamines
Neutral salts
High purity water
Coefficient
none
5.1.2.2
Sensor 2
5.1.2.2.1
Cell Constant
: Enter the cell constant printed on the sensor label.