SVAN 973 User Manual
27
In the
Simple
user interface, the
Calibration
list
comprises positions enabling calibration with the use of
the sound calibrator (
By Measure.
), checking and
erasing previous calibration records (
Last Calibration
,
Calibration History
,
Clear History
), adding current
calibration results to the logger file (
Post Cal.
).
<ENT>
In the
Advanced
user interface, there is additional position in the
Calibration
list
-
Auto Calibration
, which allows you to switch on/off the auto calibration function.
Note: It is advised to perform calibration of the instrument each time before the measurements
begin. A single calibration at the start of each day is usually sufficient for most regulations.
Note: The calibration factor is always added to measurement results and measurement range
limits of the Lev. Met., 1/1 Oct., 1/3 Oct., RT60 and Dosimeter functions.
Note: The recommended factory calibration interval is 12 months for instruments to be
confident in their continuing accuracy and compliance with the international codes. Please
contact your local Svantek distributor for further details.
Note: It is possible
to perform so called “By Sensitivity” calibration by setting the calibration
factor with the use of special command
– see Appendix A.
3.2.1 Calibration
– By Measurement
To calibrate the instrument:
1. Set the calibration level (
Cal. Level
)
– see Appendix C, Chapter C.1, par.
Calibration.
2. Carefully attach the sound calibrator (SV 34B or equivalent 114 dB / 1000 Hz)
to the instrument's microphone.
Note: It is also possible to use an electro-mechanical pistonphone, which generates the signal
(ca 124
dB) or different type of sound calibrator dedicated for ½” microphones.
3.
Switch on the calibrator (if the used calibrator doesn’t have auto run function)
and wait ca 30 seconds for the tone to stabilise before starting the calibration
measurement.
4. Start the calibration measurement by pressing the
<Start>
key.
The calibration delay time is set to 3 seconds. While waiting for the start of the
measurement the
Delay
is counting down on the display.
The calibration process is divided into calibration of the Level meter and calibration of the Dosimeter.
The progress bar displays the entire progress of the calibration.
During the first calibration, the level of the calibration signal
measured by the Level Meter is displayed until the measured
result stabilizes (the maximum difference between three
consecutive 1-second LCeq results should be within 0.05
dB). The result of the calibration will then be displayed and
the next measurement for the Dosimeter will be waiting for
the start.
. . .
Содержание SV 973
Страница 157: ...SVAN 973 User Manual Appendixes 157 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 158: ...SVAN 973 User Manual Appendixes 158 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 159: ...SVAN 973 User Manual Appendixes 159 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 172: ...SVAN 973 User Manual Appendixes 172 250 Hz 315 Hz 400 Hz 500 Hz 630 Hz 800 Hz...
Страница 173: ...SVAN 973 User Manual Appendixes 173 1000 Hz 1250 Hz 1600 Hz 2000 Hz 2240 Hz 2500 Hz...
Страница 174: ...SVAN 973 User Manual Appendixes 174 2800 Hz 3150 Hz 3550 Hz 4000 Hz 4500 Hz 5000 Hz...
Страница 185: ...SVAN 973 User Manual Appendixes 185 500 Hz 1 1 octave filter 250 Hz 1 1 octave filter...
Страница 192: ...SVAN 973 User Manual Appendixes 192 LF filter according to EPA 93 F105 02 104 Low Frequency Noise Control Regulations...