4.

 

MAINTENANCE 

4.1

 

Fitting New O-rings 

Routine use will create wear on the three black O-rings located in the tip of the micropipette 
holder chuck. Occasionally, the O-rings will need to be replaced. To do this, gently remove 
the old O-rings with an appropriate tool (a bent paper clip, for example). Discard the old O-
rings and insert three new ones, taking care to ensure that they are flat against the back of 
the chuck. The new O-rings can be pushed down inside the chuck by gently screwing the 
chuck onto the knurled black aluminum pressure fitting. Take care not to damage the chuck 
during this procedure. 

4.2

 

Troubleshooting 

Problem Cause  Solution 

1. The black O-rings in the 

chuck are damaged. 

Replace the O-rings. 

2. There is an air bubble in the 

hydraulic fluid. 

Refill the system with 

hydraulic fluid, taking care not 
to introduce air bubbles. 

There is poor control by the 

injector of the material in the 

micropipette. 

3. The syringe plunger clamp 
or barrel clamp is not tight. 

Tighten the appropriate clamp 
screws 

When drawing hydraulic fluid 

into the syringe (without 
tubing attached), air bubbles 

are forming inside the syringe 
plunger. 

When filling the syringe for the 
first time, a little air is trapped 

inside the tip of the syringe, 

which is drawn into the barrel 

when the plunger is pulled 
back. 

Slowly draw hydraulic fluid 
into the barrel, then 
immediately and rapidly expel 

it back into the container. 
Repeat this until a full barrel 

of hydraulic fluid can be drawn 
without drawing air bubbles. 

 

 
 

XENOWORKS ANALOG MICROINJECTOR OPERATION MANUAL – REV. 4

Содержание XenoWorks BRI

Страница 1: ...XenoWorks BRI Analog Microinjector OPERATION MANUAL Rev 4 00 20100521 One Digital Drive Novato CA 94949 Voice 415 883 0128 Web www sutter com Fax 415 883 0572 Email info sutter com...

Страница 2: ...ENOWORKS ANALOG MICROINJECTOR OPERATION MANUAL REV ii 4 00 20100521 Q Copyright 2008 2010 Sutter Instrument Company All Rights Reserved XenoWorks is a registered trademark of Sutter Instrument Company...

Страница 3: ......

Страница 4: ......

Страница 5: ...h pressure microinjectors Use this instrument only for microinjection purposes in conjunction with the procedures and guidelines in this manual This instrument is designed for use with capillary glass...

Страница 6: ...00 20100521 iv This page intentionally left blank XENOWORKS ANALOG MICROINJECTOR OPERATION MANUAL REV 4...

Страница 7: ...4 Maintenance 11 4 1 Fitting New O rings 11 4 2 Troubleshooting 11 Appendix A Limited Warranty 13 Appendix B Accessories and Spare Parts 15 Appendix C Technical Specifications 17 Index 19 TABLE OF FIG...

Страница 8: ...XENOWORKS ANALOG MICROINJECTOR OPERATION MANUAL REV vi 4 00 20100521 This page intentionally left blank...

Страница 9: ...e the Figure 1 1 to identify the parts listed If any items are missing or damaged contact Sutter Instrument immediately Figure 1 1 Components of the XenoWorks analog microinjector NOTE To operate the...

Страница 10: ...inge plunger that it grips in and out of the syringe barrel The sliding block is moved by rotation of the micrometer screw which can be seen running the length of the inside body of the microinjector...

Страница 11: ...E UNIT 1 2 5 Angle Adjustment This knob is used to fix the microinjector at an angle appropriate for comfortable use 1 3 Micropipette Holder The micropipette holder is attached via a length of tubing...

Страница 12: ...XENOWORKS ANALOG MICROINJECTOR OPERATION MANUAL REV 4 00 20100521 4 This page intentionally left blank...

Страница 13: ...the ferrule is pointing in the correct direction Figure 2 1 Attaching the tubing to the injector 3 Unpack the syringe and ensure the plunger is pushed all the way in NOTE If you have decided to use t...

Страница 14: ...e Volume Maximum Tubing Length 1000 l 1930 mm 500 l 965 mm 100 l 178 mm Figure 2 2 Connecting the tubing to the tip of the syringe 8 Once the hydraulic fluid has been pushed to the very tip of the tub...

Страница 15: ...lamp screw until the syringe barrel is held securely Next tighten the plunger clamp screw Take care to tighten the black syringe barrel clamp first before tightening the silver plunger screw CAUTION T...

Страница 16: ...the pipette holder 12 Adjust the tilt of the microinjector using the angle adjustment knob to achieve a comfortable angle for working and place the unit on the bench top by the micromanipulator Your X...

Страница 17: ...ad of a fluid in the microinjector skip Step 1 1 First remove the clear chuck from the front of the micropipette and turn the injector s coarse control clockwise to advance the hydraulic fluid to the...

Страница 18: ...XENOWORKS ANALOG MICROINJECTOR OPERATION MANUAL REV 10 4 00 20100521 This page intentionally left blank...

Страница 19: ...s in the chuck are damaged Replace the O rings 2 There is an air bubble in the hydraulic fluid Refill the system with hydraulic fluid taking care not to introduce air bubbles There is poor control by...

Страница 20: ...XENOWORKS ANALOG MICROINJECTOR OPERATION MANUAL REV 12 4 00 20100521 This page intentionally left blank...

Страница 21: ...trument has been operated in accordance with the instructions outlined in the instruction manual Abuse misuse or unauthorized repairs will void this warranty Limited warranty work will be performed on...

Страница 22: ...XENOWORKS ANALOG MICROINJECTOR OPERATION MANUAL REV 1 4 00 20100521 4 This page intentionally left blank...

Страница 23: ...0 l Syringe V001184 250 l Syringe V001185 500 l Syringe V001186 1 000 l Syringe BR MH Micropipette holder includes micropipette holder body 9 O rings BR AT Analog tubing kit includes 2 x 1 m ETFE tubi...

Страница 24: ...XENOWORKS ANALOG MICROINJECTOR OPERATION MANUAL REV 1 4 00 20100521 6 This page intentionally left blank...

Страница 25: ...y Prime Tech Burleigh Storage Environment Temperature 0 70 C 32 158 F Humidity 0 95 non condensing Operating Environment Temperature 3 5 35 C 38 3 95 F Humidity 0 80 non condensing Size without syring...

Страница 26: ...XENOWORKS ANALOG MICROINJECTOR OPERATION MANUAL REV 18 4 00 20100521 This page intentionally left blank...

Страница 27: ...block 2 Syringe clamp 2 D disclaimer iii G General Safety Information iii I Installation and assembly 5 M Maintenance 11 Fitting new O rings 11 N notes user 20 O Operating Instructions 9 Fitting a mic...

Страница 28: ...00 20100521 20 NOTES XENOWORKS ANALOG MICROINJECTOR OPERATION MANUAL REV 4...

Отзывы: