3.7.7.4
Programming a Filter Number (“Logical Filter”)\
The Galvanometer Values that define the maximal light output through FILTER
POSITIONS 1,2,3 and 4 have been assigned FILTER NUMBERS 1,2,3 and 4, respectively, at
the factory. It is recommended that these assignments be retained but there is no overwrite
protection. If you do edit these values, the original values can be found on the Optical
Alignment Data Sheet that came with your Lambda DG-4. Galvanometer Values, which
maximally misalign the Input and Output Galvanometers, have been assigned FILTER
NUMBER 0. When selected, this FILTER NUMBER will prevent light output and serve as a
“logical shutter”.
The EDIT A FILTER DEFINITION mode is the operational mode from which you assign
paired Galvanometer Values a FILTER NUMBER. It is selected by pressing the MODE key
followed by the 7 key to get to the EDIT A FILTER DEFINITION menu. Alternatively, you
can “page” through the mode selection menus by pressing MODE. The following menu will
be displayed:
Press 1 three times to scroll through the NEXT TWO PAGES to see the desired option or
press the 0 key to abort the routine and return to the MAIN MENU.
Press 7 to view the EDIT A FILTER DEFINITION menu.
Enter the desired FILTER NUMBER (0-15) as described in the MOVE ON COMMAND
section; for one digit values, press the desired numeric key followed by ENTER or press the 0
key then the desired numeric key; for two digit values press the numeric keys only. Pressing
the MODE key will abort the EDIT routine, return you to the first MODE SELECTION
menu and retain the current FILTER NUMBER settings. If you have entered a FILTER
NUMBER value (e.g., 4), the following edit screen will appear:
0=EXIT, 1=NEXT PAGE, 2 = TURBO BLANKING
3 = DISPLAY THE FIRST 4 FILTER VALUES
0=EXIT, 1=NEXT PAGE. 4= MOVE ON SYNC.
PULSE, 5 FOR MOVE ON STROBE PULSE
0=EXIT, 1=NEXT PAGE. PRESS 6 FOR MOVE ON
SYNC, INHIBITED BY THE STROBE LIN
0=EXIT, 1=NEXT PAGE. 7 = EDIT A FILTER
DEFINITION, 8 = SETUP THE RING BUFFER
WHICH FILTER DO YOU WISH TO EDIT OR
CREATE?
Содержание Lambda DG-4PLUS
Страница 3: ......
Страница 4: ......
Страница 16: ...LAMBDA DG 4 PLUS DG 5 PLUS OPERATION MANUAL REV 3 05B 20131009 4 This page intentionally blank...
Страница 24: ...LAMBDA DG 4 PLUS DG 5 PLUS OPERATION MANUAL REV 3 05B 20131009 12 This page intentionally blank...
Страница 46: ...LAMBDA DG 4 PLUS DG 5 PLUS OPERATION MANUAL REV 3 05B 20131009 34 This page intentionally blank...
Страница 66: ...LAMBDA DG 4 PLUS DG 5 PLUS OPERATION MANUAL REV 3 05B 20131009 54 This page intentionally blank...
Страница 68: ...LAMBDA DG 4 PLUS DG 5 PLUS OPERATION MANUAL REV 3 05B 20131009 56 This page intentionally blank...
Страница 78: ...LAMBDA DG 4 PLUS DG 5 PLUS OPERATION MANUAL REV 3 05B 20131009 66 This page intentionally blank...
Страница 82: ...LAMBDA DG 4 PLUS DG 5 PLUS OPERATION MANUAL REV 3 05B 20131009 70 NOTES...