25
APPENDIX C.
FUSE REPLACEMENT
In the event that the controller fails to power up when the power switch is turned on, check
the line power fuses to see if either or both have blown. The fuses are located in the fuse
holder on the power entry module on the back of the controller. To remove the fuse holder
first unplug the power cord from the power entry module. Press down on the lever that is
located just above the fuse holder and below the power receptacle, and pry the holder straight
out of the power entry module.
Figure 4-1. Power entry module and fuse location on the rear panel of the Lambda HPX-L5 controller.
The fuse holder holds two fuses. Both fuses are of the same type and rating. If either fuse is
blown, it is recommended that both fuses be replaced.
The type and rating of both fuses are as follows:
5 x 20 mm glass tube, Time Delay (IEC 60127-2, Sheet III)
T2A, 250V
(Examples: Bussmann GDC-2A. GMC-2A or S506-2-R (RoHS); or
Littelfuse 218 200 or 218 200P (RoHS))
LAMBDA HPX-L5 OPERATION MANUAL – REV. 1.03 (20161130)
Содержание Lambda HPX-L5
Страница 3: ......
Страница 4: ......
Страница 16: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 14 This page intentionally blank...
Страница 20: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 18 This page intentionally blank...
Страница 22: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 20 This page intentionally blank...
Страница 24: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 22 This page intentionally blank...
Страница 26: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 24 This page intentionally blank...
Страница 28: ...LAMBDA HPX L5 OPERATION MANUAL REV 1 03 20161130 26 This page intentionally blank...