42
Electromagnetic Compatibility
EN
11
Electromagnetic Compatibility
Precautionary measures
Medical devices are subject to special safety and protective measures concerning
electromagnetic compatibility (hereafter abbreviated as EMC).
This device is to be used only for the purposes described in the manual and has
to be installed, set up, and operated in compliance with the EMC notes and in-
structions.
CAUTION!
Accessories
To ensure compliance with the requirements of IEC 60601-1-2 in the current ver-
sion, the device AirSeal® iFS must be used only with the accessories listed in
chapter 14 "Accessories".
In order to ensure the basic safety and the essential performance in relation to
electromagnetic interference over the lifetime of the device, the AirSeal® iFS
must be restarted after 24 hours to allow the self test to be repeated. Further-
more, the service interval described in chapter 9.2 "Maintenance Intervals" must
also be adhered to.
CAUTION!
Continuous operation
After 24 hours of continuous operation, a device self-test must be carried out.
• Switch device off and on again.
11.1
Impact of Mobile and Portable HF Communication Devices
The emission of high frequency energy by mobile communication devices may
impact the function of the electrical medical device. Operating such devices (e.g.,
cell phones, GSM phones) in the proximity of the electrical medical device is pro-
hibited.
11.2
Electrical Connections
ESD (Electrostatic Discharge) precautionary
measures
The following are ESD precautionary measures:
• Apply potential equalization (PE), if available on your equipment, to all devices
to be connected.
• Use only the listed equipment and accessories.
Hospital employees should be informed about and trained in ESD precautionary
measures.
Содержание AirSeal iFS
Страница 1: ...EN Instructions for use...
Страница 4: ......
Страница 6: ...2 EN 15 AirSeal iFS Warranty 52 16 Test Log 53 16 1 Test Log 53 16 2 Return Form 54 Index 55...
Страница 20: ...16 Initial Device Startup EN House gas supply pressure OK House gas supply pressure too low...
Страница 57: ...Test Log 53 EN 16 Test Log 16 1 Test Log Date Results Comment Signature...
Страница 60: ...W55 186 EN AA...