34
User Menu
EN
8.1
Setting First Nominal Pressure
In the user menu, tap the
Starting Pressure
key to access the setting.
Select the desired
Starting Pressure
for the currently selected insufflation mode.
Press the
º
or
»
keys to increase or decrease the nominal pressure.
8.2
Gas Flow Rates
In the user menu, tap the
Gas Flow Rates
key to access the selection. You can se-
lect one of three gas flow rates for each insufflation mode.
Press the
º
or
»
keys to increase or decrease the gas flow.
8.3
Setting Smoke Evacuation Level
In the user menu, tap the
Smoke Evacuation Level
key to access the selection. For
AirSeal Mode a pre-setting can be chosen.
8.4
Setting General Volume
In the user menu, tap the
Audio/Display
key to access the signal volume selec-
tion.
You can select one of four volume levels for the warning signals as well as the
info and alert signals, all in the form of acoustic signals. The selection/setting ap-
plies to all insufflation operating modes.
Press the
º
or
»
key to set the desired volume.
8.5
Setting Brightness
In the user menu, tap the
Audio/Display
key to access the brightness selection.
Use the
Brightness
option to adjust the screen display to the ambient light con-
ditions within the operating room.
Insufflation operating mode Factory setting
Range
AirSeal Mode
15 mmHg
5-15 mmHg
Smoke Evacuation Mode
15 mmHg
5-15 mmHg
Standard Insufflation Mode
15 mmHg
5-15 mmHg
Insufflation operating mode
Range
AirSeal Mode (Initial Insufflation)
Gas flow between 1 and 40 l/min
Smoke Evacuation Mode
Gas flow between 1 and 40 l/min
Standard Insufflation Mode
Gas flow between 1 and 40 l/min
AirSeal Mode
LOW
HIGH
Warning signals
Alert signals
Info signals
I (Low)
I (Low)
I (Low)
II (Medium)
II (Medium)
II (Medium)
III (High)
III (High)
III (High)
IIII (Very High)
IIII (Very High)
IIII (Very High)
Содержание AirSeal iFS
Страница 1: ...EN Instructions for use...
Страница 4: ......
Страница 6: ...2 EN 15 AirSeal iFS Warranty 52 16 Test Log 53 16 1 Test Log 53 16 2 Return Form 54 Index 55...
Страница 20: ...16 Initial Device Startup EN House gas supply pressure OK House gas supply pressure too low...
Страница 57: ...Test Log 53 EN 16 Test Log 16 1 Test Log Date Results Comment Signature...
Страница 60: ...W55 186 EN AA...