20
Operating the Device
EN
After pressing the operating mode key, an informational window opens for the
required tube (only for
AirSeal
and
Smoke Evacuation Mode
). Insert the tube set
by positioning the filter canister into the AirSeal® iFS and pulling down the lever
(
AirSeal and Smoke Evacuation Mode)
or by pushing the open side of the tube set
over the barbed connector
(Standard Insufflation Mode)
.
5.5.3
Starting/Stopping Insufflation
Start insufflation:
Press
START
to start insufflation.
Activated insufflation:
The status line depicts
Initial Insufflation
.
The
Initial Insufflation
phase remains
in operation until the set nominal pressure has been reached through the insuf-
flation for the first time.
CAUTION!
For the safety of the patient please fill the tube set with CO
2
gas prior to begin-
ning the insufflation by activating the insufflation for a few seconds and then
turning it off again before introducing the insufflation instrument to the cavity
and beginning the surgery.
Stop insufflation:
Press
STOP
to stop insufflation.
SurgiQuest AirSeal® Blunt
Tip Obturator (hereafter
referred to as Blunt Tip Obtu-
rator)
SurgiQuest AirSeal® Cannula
(hereafter referred to as
AirSeal® Cannula)
AirSeal® Cannula and Low
Profile Obturator with Blade-
less Optical Tip (hereafter
referred to as AirSeal® Access
Port)
Содержание AirSeal iFS
Страница 1: ...EN Instructions for use...
Страница 4: ......
Страница 6: ...2 EN 15 AirSeal iFS Warranty 52 16 Test Log 53 16 1 Test Log 53 16 2 Return Form 54 Index 55...
Страница 20: ...16 Initial Device Startup EN House gas supply pressure OK House gas supply pressure too low...
Страница 57: ...Test Log 53 EN 16 Test Log 16 1 Test Log Date Results Comment Signature...
Страница 60: ...W55 186 EN AA...