Initial Device Startup
13
EN
CAUTION!
Ventilation of the device
Avoid device overheating. Ensure free air circulation especially to the bottom
and rear of the device (rear panel distance of at least 10 cm).
CAUTION!
Position the device in such a way that it is easy to operate and switch off.
CAUTION!
Place the device outside the sterile field.
CAUTION!
Position of the user
To avoid a malfunction, the user must be positioned correctly
• within a display viewing angle of ±50° to operate the device
• up to 2 m/6.5 ft from the device front for monitoring the actual values
Mains connection
CAUTION!
Check to make sure the available mains voltage matches the data listed on the
type label attached to the back of the device. Incorrect voltage can cause errors
and malfunctions and may destroy the device.
Make sure the connection data and technical specifications of the power supply
comply with DIN VDE or national requirements. The mains power supply cable
must be plugged into a properly installed safety wall plug (see DIN VDE 0100-
710). Read the device label located in rear of device (type plate) to determine the
operating voltage of the device.
Grounding contact
The power connection must be equipped with a grounding contact. Use the orig-
inal power cable (if included in scope of delivery) to establish a connection be-
tween the mains wall socket and the non-heating device plug located in the rear
of the device.
Only for U.S. operators
Only use a certified (UL-listed), removable mains connection cable, type SJT, min-
imal 18 AWG, 3 leads. The plug connectors must comply with NEMA 5-15 and IEC
16320-C13. Grounding will only be reliable if the equipment is connected to a
corresponding hospital grade socket.
Potential equalization
The potential equalization denotes a highly electrically conductive connection,
minimizing the different electrical potentials according to requirements of IEC
60601-1 in the respectively valid version. The installation must be according to
the relevant local safety regulations.
4.2
Gas Connection
WARNING!
Medically pure CO
2
Make sure to use only medically pure CO
2
. Other gases (e.g., helium, N
2
O, ar-
gon), mixtures of gases, gases with entrapped liquids, or polluted gases must not
be used with this device.
Use a high-pressure hose to connect a CO
2
gas cylinder to the rear gas inlet con-
Содержание AirSeal iFS
Страница 1: ...EN Instructions for use...
Страница 4: ......
Страница 6: ...2 EN 15 AirSeal iFS Warranty 52 16 Test Log 53 16 1 Test Log 53 16 2 Return Form 54 Index 55...
Страница 20: ...16 Initial Device Startup EN House gas supply pressure OK House gas supply pressure too low...
Страница 57: ...Test Log 53 EN 16 Test Log 16 1 Test Log Date Results Comment Signature...
Страница 60: ...W55 186 EN AA...