12
Initial Device Startup
EN
4
Initial Device Startup
The operation of the device AirSeal® iFS is reserved for medical staff with the rel-
evant professional qualifications trained to use the device.
Delivery inspection
Always check all parts and accessories of the device immediately after receiving
the shipment. The manufacturer considers only replacement claims that have
been immediately submitted or reported to a sales representative or an autho-
rized agent.
Setting up the device
Place the device on a flat surface and install in dry environment. The ambient
temperature and humidity must meet the requirements mentioned in chapter
13 "Technical Data".
Ensure good ventilation of the device. The vents are located on the underside and
on the back (see 5.2 "Device Rear" and 5.3 "Bottom of the Device").
NOTE!
Locations
The device is only to be used in a professional facility healthcare enviroment.
WARNING!
Not explosion-proof
The device is not explosion-proof. Do not operate the device in the vicinity of ex-
plosive anesthetic gases and not in the vicinity of oxygen-enriched environ-
ments.
4.1
Device Setup and Connection
Setting up
Place the device on a flat surface free of vibration located in a dry environment.
The ambient temperature and humidity must meet the requirements mentioned
in chapter 13 "Technical Data".
WARNING!
Use only parts and/or devices from ME systems (see chapter 11 "Electromagnetic
Compatibility") in patient environments in compliance with the standard
IEC60601-1 in the respective currently valid version.
CAUTION!
ME Device in Rack
If it should be necessary to operate the device close to stacked with other devic-
es, the ME device or ME system (see chapter 11 "Electromagnetic Compatibility")
should be monitored to ensure it works properly as configured.
CAUTION!
The medical electrical (ME) device is suitable for integration in ME equipment
systems (see chapter 11 "Electromagnetic Compatibility"). Operation of the ME
device in vicinity of non-ME devices may result in voiding the intended use of the
ME device.
CAUTION!
Equipment should be positioned such that power cord can be easily disconnect-
ed.
Содержание AirSeal iFS
Страница 1: ...EN Instructions for use...
Страница 4: ......
Страница 6: ...2 EN 15 AirSeal iFS Warranty 52 16 Test Log 53 16 1 Test Log 53 16 2 Return Form 54 Index 55...
Страница 20: ...16 Initial Device Startup EN House gas supply pressure OK House gas supply pressure too low...
Страница 57: ...Test Log 53 EN 16 Test Log 16 1 Test Log Date Results Comment Signature...
Страница 60: ...W55 186 EN AA...