PickUp-ToF - Manual
4 Device Layout & Operation
4.1 PickUp-ToF – Operation Voltages
There are 2 defined lens settings available for the PickUp-ToF to be operated in two different operation
modes. The first mode has a higher, the second mode a lower operation voltage for the collector. These two
settings are given below in Table 2.
Lens Name
HV Channel/ cable label
OperationVoltage (Mode 1)
Operation Voltage (Mode 2)
Trans1a
HV3a/ HV3(1)
400V
400V
Collector
HV2
200V
50V
ToF
HV1
50V
50V
Trans1b
HV4a/ HV4(1)
400V
400V
DetFoc
HV6
250V (5*ToF Voltage)
250V (5*ToF Voltage)
Trans2a
HV3b/ HV3(2)
800V
800V
Trans2b
HV4b/ HV4(2)
800V
800V
Focus
HV5
2000V
1500V
Table 2: PickUp-ToF Operation Voltages
4.2 PickUp-ToF – Schematic Lens Setup and Trajectory Simulation
Figure 2 shows schematic lens setup of the PickUp-ToF as well as the result of trajectory simulation for the
operation mode 1. Arrows with the same color mark identical potentials.
10