Chapter 3 Display Panel
www.supmea.com
- 9 -
3.3 LED Indicator
WASH: Cleaning control relay status indicator
RELAY A: Alarm control relay A status indicator
RELAY B: Alarm control relay B status indicator
3.4 Default setting
Password
:
0000
Mode
:
pH
Calibration type: 3-point calibration, 4.00 / 6.86 / 9.18
Temperature type: MTC 25.0
℃
Relay A: ON, Hi, high point 11.00pH, return difference 1.00pH
Relay B: ON, Lo, low point 3.00pH, return difference 1.00pH
Cleaning relay: OFF
Communication: baud rate 9600, no parity, 1 stop bit, floating point format 4321,
address 1
Current output A: pH, 4 ~ 20mA, -2.00 ~ 16.00pH
Current output B: T, 4 ~ 20mA, -10.0 ~ 130.0
℃
Date and time: 0: 0: 0: 0 on January 1, 2018
Filter coefficient: 0
Backlight brightness: 20
Recording interval: 15 minutes