Chapter 5 Programmable Output
-
www.supmea.com
-20-
Chapter 5 Programmable Output
5.1 Split Output Function (n/m)
Through split output, voltage, current and thermocouple signals
can be split and output in n/m times.
Output value = (main set value) x (n/m)
Press
or
to set the main set value for output.
If there is no wave form, press the programming
button
to access the split output mode. The
n/m menu is displayed.
Press the programming button
to set the m
value (value range: 1 to 20).
Press
or
to set the n value (value range: 0
to m). The n/m output value can be obtained base on
the n value.
Press the yellow source button
, and the value of
on the screen will turn from
to
,
and the output is started.
Press the yellow source button
to stop the
output.
Press the programming button
to exit the split
output mode.