Quick Guide to Display Messages
Symbol:
Definition:
*
Calibration has been adjusted from factory setting for the
grain being displayed
Tempe+
Temperature above limit
Temperature --
Temperature below limit
Hi
Moisture content above upper limit
Lo
Moisture content below lower limit
Change battery
Battery needs replacing
Technical Specifications
●
Measuring range
(Moisture): 5 to 45% Depending on crop type
●
Operating range
(Temperature): 0 to 50 ºC
●
Resolution
: 0.1% moisture
●
Temperature compensation
: Automatic
●
Weight including case
: 1.6 kg
●
Dimension
: 13x21x8 cm
●
Battery
: 9 V Alkaline
●
Display
: 2 line LCD with backlight
●
Accuracy
: Better than +/- 0.5% in normal moisture range,by using average and
calibration adjustments
Troubleshooting
➢
The meter does not turn on by pressing the ON/TEST button
Change the battery. It is a good idea to always have a new battery in your carrying-case
➢
The meter shows Lo<xx% in the display
The moisture content of the sample in the measuring cell is below the measuring range
➢
The meter shows Hi>xx% in the display
The moisture content of the sample in the measuring cell is over the the measuring range
Or
Ver 1.1.0
Page 11/12