OXYGEN CONCENTRATORS

M6000 & M6100

Operators Manual

Supera , LLC Copyright 2021

For Veterinary Use Only

Model M6000

*Shown with optional vaporizer

Model M6100

*Shown with optional vaporizer

Содержание Pureline M6000

Страница 1: ...OXYGEN CONCENTRATORS M6000 M6100 Operators Manual Supera LLC Copyright 2021 For Veterinary Use Only Model M6000 Shown with optional vaporizer Model M6100 Shown with optional vaporizer...

Страница 2: ......

Страница 3: ...sales representative right away for help Our goal is to make your new anesthesia machine as easy to use and care for as possible This device is meant to be operated under the normal supervision and co...

Страница 4: ...supply hose Oxygen Concentrator Features Operation Instructions Maintenance Features Accessories Safety Pop Off Valve Rebreathing circuits Non rebreathing circuits I V Pole Optional Items Waste gas ev...

Страница 5: ...ghten both screws with the 3 16 allen wrench provided VERY tightly NOTE THIS REQUIRES TWO PEOPLE TO SAFELY ASSEM BLE 1 Carefully unpack the top assembly and frame from the box 2 Align the top to the f...

Страница 6: ...1 Remove the oxygen concentrator from the box and guide the four threaded post into the frame 2 Install and tighten the nuts DO NOT OVER TIGHTEN THE NUTS SNUG FIT IS ALL THAT IS NEEDED...

Страница 7: ...1 Attach the green hose to the threaded fitting DO NOT OVER TIGHTEN...

Страница 8: ...Model M6100 non rebreathing anesthesia machine Includes a stainless steel storage tray and 3 8 post bulkheads for mounting accessories i e I V poles multi port manifolds etc...

Страница 9: ...F power switched off Class II protection Do not expose to open flames Do not use oil or grease Technical information Consult the accompanying documents Keep in the vertical position Fragile handle wit...

Страница 10: ...entire manual should be allowed to operate the O2 concentrator The CAUTIONS below indicate a potentially hazardous situation If conditions are not avoided a situation could occur that results in prop...

Страница 11: ...en Concentrator is packaged to protect the device from damage while being transported and stored Check for damage to the packaging After device is removed from the package inspect for damage If damage...

Страница 12: ...2 1 Front panel Fig 2 1 1 I O ON OFF switch 2 Indicator lights 3 Oxygen product outlet 4 Circuit breaker 5 Pre set flow meter non adjustable...

Страница 13: ...pproximately one minute after the device is stopped 1 Filter Silencer 2 Cabinet filter 3 Ventilation grill Note Shown with grate removed Fig 4 1 4 CLEANING MAINTENANCE Only the outside of the concentr...

Страница 14: ...oduction Thecolumns contain a molecular sieve whose function is to adsorb the nitrogen and thus allow oxygen to pass The oxygen enriched product is then directed through a pressure reducing valve cont...

Страница 15: ...fitted on the compressor outlet and is calibrated to 2 7 bar 40 psig 5 4 3 Indicators The green indicator light Fig 5 1 indi catesthatpowerisappliedtothedevice When first turned on the indicator will...

Страница 16: ...4 Maintenance of the Device Alarms No special maintenance is required The alarm set point is factory set and the setting cannotbeadjusted Allmodelsaresetat84 The equipment supplier verifies that the...

Страница 17: ...compressor compartment Environmental limit conditions The performances of the device especially the oxygen concentra tion are quoted at 21o C 70o F and one atmosphere They may change with temperature...

Страница 18: ...Pneumatic connection broken or other pressure problem Internal electrical fault Pneumatic circuit fault Dirty Filters blockage Fan is not working Observations The I O ON OFF button is in the I ON posi...

Страница 19: ...Anesthesia Machine Operation Optional Accessories...

Страница 20: ...breathing circuit is partially closed but will leak at 0 5 cm H2O This will not cause injury to the patient however depressing the push button valve to ventilate the patient would allow excessive pres...

Страница 21: ...4 Add pressure to the system until the manometer reads 30 cmH20 Hold the pressure for ten seconds If it doesn t drop faster than 1 cm H2O per second the machine is sealed enough for use If it does dro...

Страница 22: ...ith the switch in the AUX position the supply of oxygen to the anesthesia machine comes from E tank on the back of the machine or another source using the E tank adapter p n OC6050 To use the E tank a...

Страница 23: ...Accessories Shown with optional universal control arm for Bain circuits P N MA2013 safety pop off valve shown P N MA2012 Regular pop off valve not shown I V Pole included...

Страница 24: ...Optional F Air evacuation kit P N EVC555 Bag circuit organizer Features Accessories...

Страница 25: ...face valve P N EVC630 for use with fan based evacua tion systems like our p n EVC3000 Optional waste gas interface valve P N EVC629 for use with vacuum pump based evacuation systems like our p n EVC31...

Страница 26: ...connector from the non rebreathing system into the com mon gas outlet of the anesthesia machine 3 Connect the blue exhaust port on the bag of the non rebreathing system to the 19mm blue evacuation tub...

Страница 27: ...l connector at the patient end elimi nates kinking of the tubing This circuit attaches to the Inhalation and Exhalation ports on the rebreathing head CO2 Absorber P N CIR529 Adult blue inner tube Reco...

Страница 28: ...res sure test Contact Supera for a list of certified service companies Daily Inspection Leak Test Before each use check for pressure leaks in the machine Also make certain the waste gas evacuation sys...

Страница 29: ...lation and or Installation Operation manual applicable for the product 3 products considered to be of a consumable nature such as rubber or plastic goods 4 accessories or parts not manufactured by Sup...

Страница 30: ...company s terminal where they are entirely responsible 5 Open and inspect your product as soon as possible DO NOT WAIT 6 If you find damage take as many photographs of everything as soon as you can a...

Страница 31: ...ATION Date purchased Purchased from Machine serial number Vaporizer serial number Machine service information Service Date Service information Service Date Service information Service Date Service inf...

Страница 32: ...503 723 5068 or 877 620 1500 toll free E Mail info superavet com 1220 NW Noble Drive Estacada OR 97023 www SuperaVet com OXYGEN CONCENTRATORS...