Specifications
Method of
Measurement:
Oscillometry with step deflation
Blood Pressure Range:
25-260 mmHg (max inflate 280mmHg)
Heart Rate Range:
40-200 bpm
Accuracy:
Clinically validated to ESH International Protocol, BHS
(A/A), ANSI/AAMI (SP10)
International
Standards:
EN 60601-1, EN 60601-2-30,
EN 60601-1-2 (EMC),
EN 1060-1, EN 1060-3, “Non-Invasive
Sphygmomanometers -General Requirements
& Supplementary Requirements For Electro-
Mechanical BP Measuring Systems”, AAMI
SP10 ES1 category C’ (battery powered)
Operating Conditions:
10°C (50°F) to 50°C (122°F)
20-95% RH non-condensing
Power:
Two “AA” alkaline batteries or high capacity
rechargeable batteries (NiMH)
Data Memory:
Flash memory stores up to 250 readings
Calibration:
Minimally, once every two years
Safety Systems:
Maximum inflation pressure limited to 300 mmHg;
Auto safety release valve for power failure; Maximum
BP measurement time limited to less than 140 seconds
Sampling Periods:
3 independently programmable periods (5, 10, 15,
20, 30, 45, 60, 90 and 120 minutes)
Size:
Approximately 120 x 70 x 32 mm
Weight:
Approx 284g, including batteries
Storage Conditions:
-20°c to 70°c, 15%-95% RH non-condensing
Data Connection:
USB (RS-232 option)
7