NOTE: IR temperature readings provided by this device are closely aligned to core temperatures rather than an axillary
or rectal readings. Take this into account when reviewing readings.
In order to change the measurement units from °C to °F or vice-versa, ensure that the thermometer is on by pressing
the START button. Then press the MODE and MEM buttons on the thermometer simultaneously for 3-4 seconds.
The SunTech Touchless Thermometer can also be used to measure the temperature of an object, such as a vessel
containing liquid, or food. To use Object Mode, ensure that the thermometer is turned on by pressing the START
button. Press the Mode Button briefly. The Object Mode icon
will then appear in the device display. Place the
thermometer 2-3 cm from the desired measuring point on the object and press the START button. Hold thermometer
in place until you hear a long beep. The temperature will be displayed on the screen.
NOTE: Remember that surface temperature and inner temperature of an object can vary greatly.
*Touchless Thermometry module is not available in all countries. Contact SunTech to find out if this feature is available
in your area
9. Using SunTech CT40 EMR and Memory Functions
Section 4 “Good to Know Before You Begin”
provides some basic information about the SunTech CT40
’s
EMR and
memory functions. Here are more details about how these work.
Memory Mode
When you power up the unit, the memory identifier appears at the top of the display beside the Memory Icon
.
This is the number that will be associated with the next set of measurement results that are sent to the device
memory.
Press the Memory Button
on the front of the device to save results to memory. The Memory Icon, check
mark icon and all values that are displayed on the screen (at the time the button is pressed) will flash before the
results are saved. The following values can be saved to memory.
Parameter Name
Field Type
Patient ID
Alphanumeric Text
Time Stamp
Numeric, HH:MM:SS
Date Stamp
Numeric, MM.DD.YYYY or DD.MM.YYYY
Memory Location
Numeric
Systolic BP
Numeric
Diastolic BP
Numeric
BP Type
Text: AVG or SM