930482 Rev. A

X .   O p e r a t i n g   G u i d e

33

2. To Program Your Chair: 

a. Turn 

off

power to the chair. 

b. Insert Qtronix Programmer plug (A) into the

Programmer/Charger socket (B) on the front of
the joystick.

c. Turn on power to the chair; program as needed. 
d. Remove plug and cycle ON/OFF switch to drive

chair.

NOTE–

Review the Qtronix Programmer Owner’s Manual
for more details on how to program your chair.

NOTE–

Program settings that are not matched for the
capabilities of the rider can result in serious
injury. Do not alter settings without the advice
of your health care professional.

C. THERMAL ROLL-BACK 

Your chair has a thermal roll-back circuit. 
This protects the controller from damage due to over-
heating. In extreme conditions (such as repetitive hill
climbing) the circuit will decrease the power to your
motors. This allows the chair to operate at a reduced
speed. When the controller cools, the chair will return
to normal speed.

D. CIRCUIT BREAKERS 

1.  Notes: 

Your Quickie V-100 has two battery boxes, with a
circuit breaker on each box.

•  In the unlikely event of a short circuit or heavy

overload, all power to your chair will be shut off. 

•  To reset your chair, depress the circuit breaker

button(s) on the side of the battery boxes. A few
minutes wait is required before the circuit break-
er(s) will reset. 

2. Repeated Shutdown: 

If the chair continues to shut down after 
resetting, have it serviced by a supplier.

X .   O p e r a t i n g   G u i d e

930482 Rev. A

32

A. PERFORMANCE CONTROL SETTINGS

1. It is vital to match control settings to your level

of function and ability. 

2. Consult your health care professional and your sup-

plier to select the best control settings for you. 

3. Check and adjust the settings every six to twelve

months (or more often, if needed). 

4. Adjust the control settings immediately if you

notice any change in your ability to: 
•Control the joystick. 
•Hold your torso erect. 
•Avoid running into objects.

5. Control Settings are adjusted through the use of

the Quickie QTRONIX Programming Pad. See your
supplier if you do not own a Quickie QTRONIX
Programming Pad.

B. QUICKIE QTRONIX PROGRAMMING PAD (OPTIONAL)

Program settings beyond the ability of the rider
can result in serious injury. Consult your health
care advisor before you alter settings.

1. Notes: 

The Qtronix Programmer lets you adjust the control
settings of the Quickie V-100. You can customize a
program for yourself or select the 
standard program. 

• acceleration rate
• deceleration rate
• turn acceleration
• turn deceleration
• forward speed (max & min)
• reverse speed (max & min)
• turn speed (max & min)
• steer correct
• 4-axis joystick throw
• sleep timer
• read timer
• profiles

QTRONIX Programming Pad

A

A

B

All manuals and user guides at all-guides.com

Содержание Quickie V-100

Страница 1: ...l Warranty SUPPLIER THIS MANUAL MUST BE GIVEN TO THE RIDER OF THIS WHEELCHAIR RIDER BEFORE USING THIS WHEELCHAIR READ THIS ENTIRE MANUAL AND SAVE FOR FUTURE REFERENCE All manuals and user guides at al...

Страница 2: ...ptions to increase your use and enjoyment of this wheelchair If you lose your war ranty card call or write and we will gladly send you a new one FOR ANSWERS TO YOUR QUESTIONS Your authorized supplier...

Страница 3: ...IAGRAM 46 XIV SUNRISE LIMITED WARRANTY 47 I I T a b l e o f C o n t e n t s 930482 Rev A 4 I INTRODUCTION 3 II TABLE OF CONTENTS 4 III YOUR CHAIR AND ITS PARTS 6 IV NOTICE READ BEFORE USE 8 V EMI ELEC...

Страница 4: ...grests angle adjustable extended and platform flip up footplates toe loops exten sion tubes Casters Standard 8 pneumatic Option 8 airless insert Armrests Standard height adjustable with standard or fu...

Страница 5: ...ation devices NOTE These devices can transmit signals while they are on even if not in use I V N o t i c e R e a d B e f o r e U s e 930482 Rev A 8 A CHOOSE THE RIGHT CHAIR SAFETY OPTIONS Sunrise prov...

Страница 6: ...rn or climb or descend at an angle a fall or tip over is likely 4 Remind the rider to lean back when you tilt the chair backward 5 Lower the chair slowly Do not let the chair drop to the pavement or g...

Страница 7: ...chair on a slope or ramp if there is snow ice water or oil film present When in doubt have someone help you 3 When not in use keep your chair in a clean dry place V I G e n e r a l W a r n i n g s 93...

Страница 8: ...This will reduce the risk that you will miss the seat or fall V I G e n e r a l W a r n i n g s 930482 Rev A 14 J TERRAIN 1 This chair is designed for use on firm even surfaces such as concrete asphal...

Страница 9: ...use a fall or tip over 4 A drop off at the bottom of a slope A drop off of as small as 3 4 inch can stop a front caster and cause the chair to tip forward V I G e n e r a l W a r n i n g s 930482 Rev...

Страница 10: ...tom of the platform These may cause a fall or tip over When in doubt have someone help you 3 Always position the rider securely in the chair to help prevent falls while on a lift 4 Avoid moving forwar...

Страница 11: ...the relief of pressure 2 If you suffer from pressure sores or if you are at risk that they will occur you may need a special seat system or a device to control your posture Consult your doctor nurse...

Страница 12: ...tose or agitated K PUSH HANDLES 1 Push handles provide secure points for an attendant to propel and control the chair This helps to prevent a fall or tip over 2 Check to make sure push handle grips wi...

Страница 13: ...s balance point This will lift the front casters off the curb or step 6 Keep the chair in balance and take small steps backward Be sure to look where you are going Turn the chair around and gently low...

Страница 14: ...ill operate smoothly Check to see that all components work properly If you detect a problem be sure to correct it before use I X S e t U p A d j u s t m e n t U s e 930482 Rev A 26 NOTES 1 Work Surfac...

Страница 15: ...s e 930482 Rev A 28 A BATTERY REMOVAL To remove the batteries for transporting or servicing 1 Remove Batteries a Unplug batteries from controller b Unfasten the battery hold down straps A c Remove ba...

Страница 16: ...tails on how to contact your authorized sup plier or Sunrise customer service I X S e t U p A d j u s t m e n t U s e 930482 Rev A 30 5 Adjusting Armrest Receiver Fit To tighten or loosen the fit of t...

Страница 17: ...it is required before the circuit break er s will reset 2 Repeated Shutdown If the chair continues to shut down after resetting have it serviced by a supplier X O p e r a t i n g G u i d e 930482 Rev...

Страница 18: ...cularly if the chair is occupied X O p e r a t i n g G u i d e 930482 Rev A 34 E INTEGRAL JOYSTICK The standard integral joystick controls the chair s per formance It consists of the following parts 1...

Страница 19: ...at least 15 minutes Seek medical attention immediately X I B a t t e r i e s 930482 Rev A 36 A INTRODUCTION 1 Notes Batteries supply the power for your chair They contain a finite amount of energy and...

Страница 20: ...scharged state Unused or stored batteries should be given a charge once per month 7 Charge batteries by engaging the battery charger connector A to the charger programmer port B located on the joystic...

Страница 21: ...rs Clean around motor area weekly with a slightly damp not wet cloth Wipe off or blow away any fluff dust or dirt on or around the motors NOTE You do not need to grease or oil the chair 3 Upholstery H...

Страница 22: ...bearing spacer is in place and all four bolts are secure NOTE Use a torque setting of 240 inch pounds when tightening bolts X I I M a i n t e n a n c e 930482 Rev A 42 D BATTERY MAINTENANCE 1 To preve...

Страница 23: ...a torque setting of 420 in lbs to re tighten nut Bend washer tabs up to secure nut and install plastic dome plug g Inflate tire to proper pressure level listed on tire sidewall G MOTOR BRUSHES Check t...

Страница 24: ...r altered c This warranty applies in the USA only Check with your supplier to find out if international warranties apply 4 WHAT WE WILL DO Our sole liability is to repair or replace covered parts This...

Страница 25: ...Sunrise Medical 7477 East Dry Creek Parkway Longmont Colorado 80503 USA 800 333 4000 In Canada 800 263 3390 2002 Sunrise Medical 12 02 930482 Rev A All manuals and user guides at all guides com...

Отзывы: