Sunnex TRI-STAR Service & Installation 
Manual 

 

 

 

 

 

 

Sunnex Customer Service

 

 

Sunnex, LLC 

 

+ 1 800 445 7869

 

Warranty

 

+ 1 800 445 7869

 

 

 
10. TROUBLESHOOTING

 

 

 

Please refer to the following table for troubleshooting. 

CAUTION! DISCONNECT THE LIGHT FROM THE POWER

 

SUPPLY BEFORE ATTEMPTING ANY OF THE 
ELECTRICAL CHECKS MENTIONED BELOW.

 

 

 

PROBLEM

 

POSSIBLE CAUSE

 

SOLUTION

 

No light from light head

 

Light head is on OFF 
position

 

Turn on the ON/OFF switch to 
ON position

 

No light from light head, when 
ON/OFF switch is ON

 

Bulb has not been 
installed or it has burned 
out

 

Refer to bulb replacement 
procedure in the manual

 

 

Fuse is blown

 

Inspect / replace the fuse 
following the fuse replacement 
procedure

 

 

No power to the 
transformer box

 

Check for facility power to the 
transformer box

 

 

Transformer is 
malfunctioning

 

Check for 14 V AC output at 
the transformer when it is 
under load.

 

 

Incorrect bulb installed

 

Install correct bulb

 

Bulb burns out quickly

 

Incorrect bulb installed in 
the light head

 

Install correct bulb

 

Light output is irregular in its 
shape or intensity

 

Incorrect bulb installed in 
the light head

 

Install correct bulb

 

 

Bulb/reflector  is  not 
seated  in  its  holder 
correctly

 

Adjust the bulb/reflector so it 
seats in the holder properly.

 

No light from light head even 
when the bulb and fuse are in 
good condition, light head 
flashes intermittently when light 
head or arms are moved

 

Circuit or wiring problem 
within transformer, arm 
assemblies or other part 
of lighting system

 

Check all wiring connections. If 
problem cannot be repaired, 
call Sunnex Customer Service

 

Содержание TRI-STAR TS2015M

Страница 1: ...SUNNEX MEDICAL LIGHT MOBILE TS2035M TS2015M WALL TS2035W TS2015W TRI STAR TM INSTALLATION AND SERVICE MANUAL...

Страница 2: ...TALLATION OF TRISTAR WALL 5 6 ABOUT YOUR TRI STAR 7 7 CARE FOR YOUR TRI STAR LIGHT 8 8 SERVICE AND MAINTENANCE 9 9 WARRANTY 12 10 TROUBLESHOOTING 13 SAVE THIS PRODUCT MANUAL READ ALL INSTRUCTIONS BEFO...

Страница 3: ...THE SHIPMENT REPORT SHORTAGES AND OR ERRORS TO SUNNEX OR YOUR DEALER S CUSTOMER SERVICE DEPARTMENT WITHIN FIVE 5 WORKING DAYS ALL FREIGHT CLAIMS BY EITHER THE DEALER OR CONSIGNEE SHOULD BE FILED WITH...

Страница 4: ...n Bulb 12V 3x35W 10 20 W MODEL Halogen Bulb 12V 3x20W 10 Color temp 3 000K Medical Standard Class II Listing Weight 25 lbs Mobile 20 4 lbs Wall This product is rated as a Class II device mode of opera...

Страница 5: ...LE Remove the upright column and the lamp head assembly from the shipping carton Insert the threaded end of the upright column and screw the upright column into the base Mount the screw with washer to...

Страница 6: ...d These should be purchased to meet local codes and to ensure they are appropriate for your particular mounting surface The length of the board has been chosen so that two wall studs may be engaged du...

Страница 7: ...4 Plug the lamp head cord into the outlet in the transformer enclosure 5 Plug the three prong power cord from the transformer enclosure into the wall outlet 6 ABOUT YOUR TRI STAR TRI STAR LAMP HEAD T...

Страница 8: ...n allows you one hand control of the height of your lamp Simply squeeze the control at the top of the column and you can raise or lower the column easily The lamp head positioning arm connects with an...

Страница 9: ...FLUIDS TO THE INSIDE OF THE LAMP HEAD HOUSING The metal components of your light are powder coated for long life They too can be cleaned by using a damp cloth of mild soap and water solution mineral...

Страница 10: ...humbs and place two index fingers at flat part of spring two red lines in figure above push the spring away from you pulling it out from the slot Keep the other two ends of spring inside its holes 6 R...

Страница 11: ...cifications USA 110V Buss Fuse MDA 2 250 Volt OR Interpower 813MDA 2 250V 2 Amp Fuse MUST USE A SLOW BLOW FUSE Europe 230V Littlefuse T 1 25A 218 250V 1 25 Amp Fuse MUST USE A SLOW BLOW FUSE 6 Follow...

Страница 12: ...ICULAR NEEDS IF IMPLIED WARRANTIES MAY NOT BE DISCLAIMED UNDER APPLICABLE LAW THEN ANY IMPLIED WARRANTIES THAT MAY BE IMPOSED BY LAW ARE LIMITED IN DURATION TO THE LIMITED WARRANTY PERIOD SET FORTH HE...

Страница 13: ...Sunnex receives the Product Standard shipments to the Customer will be paid by Sunnex Sunnex assumes no responsibility for shipment delays by the carrier f If Customer requires warranty service to be...

Страница 14: ...e No power to the transformer box Check for facility power to the transformer box Transformer is malfunctioning Check for 14 V AC output at the transformer when it is under load Incorrect bulb install...

Страница 15: ...www sunnexmedical com Sunnex Inc 8001 Tower Point Drive Charlotte NC 28227 Ph 800 445 7869 Fax 888 668 1920 REV 4 September 22 2019...

Отзывы: