800-433-2797 www.Sun-Med.com
User Guide
To Use the Nebulizer with a Mask
1. Remove the mouthpiece and place in respiratory
equipment bag, in case you need it later.
2. Remove the inspiratory valve and place on the side elbow
port. (A) . To ensure a proper fit, push and turn the valve
into the side port.
3. Keep the nebulizer upright. Add the prescribed medication
through top of the reservoir tower (B).
4. Insert mask connector into the top of the reservoir tower
(C). If not using a Salter Labs aerosol mask, use the universal
mask.
5. To start treatment, turn on the air compressor switch.
Or, attach to gas source and set flowmeter 5 LPM to 8 LPM.
Check for aerosol mist in mask.
6. Apply the mask to your face.
Alternative Mask Placements
If the patient is unable to sit upright or is in a semi-fowlers
position, attach the mask to the nebulizer’s side port.
Setup A
1. Connect mask to the side port.
Setup B
1. Additional equipment is required. Gather a standard
15 mm ID / 22 mm OD adapter and a 6 inch corrugated
aerosol tubing with 22 mm ID.
2. Attach adapter to side port.
3. Attach one end of corrugate tubing to the adapter
and the other end to the aerosol mask.
Nebutech
®
8660 Reusable Nebulizer
Setup A
Setup B
B
A
C