I
NSTALLATION
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
109
SUEZ © 2022
•
Standby
— Select one of the following values to set when the Analyzer
switches out of analysis mode into standby mode:
1 mA
,
2.5 mA
,
4 mA
,
20 mA
,
22 mA
or
Hold Last
.
•
Error
— Select one of the following values to set when the Analyzer issues
an error:
1 mA
,
2.5 mA
,
4 mA
,
20 mA
,
22 mA
or
Hold Last
.
•
Warning
— Select one of the following values to set when the Analyzer
issues a warning:
1 mA
,
2.5 mA
,
4 mA
,
20 mA
,
22 mA
or
Hold Last.
4.
In the area for the analog output to configure (1, 2, or 3), specify the calibration
values:
•
Value
— Select the analog output value (TOC, TC, IC).
•
Min Value
— Enter the lowest value (ppb) for output.
•
Max Value
— Enter the highest value (ppb) for output.
•
System Protocols
— Select one of the following for running system
protocols:
Go To Standby
to output the Standby value entered in step
or
Output Values
to output actual values from protocol analyses.
To set the binary input
Enable the binary input connection for communicating remote start and stop
commands to the Analyzer.
1. On the
Settings
s
screen, select the
Hardware
tab.