I
NSTALLATION
.............................
Sievers M500e TOC Analyzer
Operation and Maintenance Manual
DLM 78510-02 EN Rev. B
108
SUEZ © 2022
6.
Repeat steps
for each alarm to configure.
7.
Click
Save
.
To set up 4-20 mA analog outputs
1. On the
Settings
s
screen, select the
Hardware
tab.
2.
On the scrolling menu, select
Analog Outputs Setup
and click
Run
. The
Analog Outputs Setup
screen appears.
3.
In the
Analog Outputs - Standby, Error, Warning Values
area, select values for
each of the following: