Figure 1

Fuse

Manual lid 

release

CAUTION: Disconnect the Microcentrifuge 

from the power outlet before cleaning. Never 

pour liquid into the unit or submerge in water

 

Maintenance & Cleaning

The Microcentrifuge is maintenance free other than occasional cleaning. There are 

no user serviceable parts inside the housing other than replacing the fuse. 
•  To replace the fuse, make sure the On/Off switch is in 

the "Off'' position, place the unit on its side and insert 

a fl at head screw driver into the fuse holder found 

on the base and turn counter-clockwise until the 

fuse is released. Replace the fuse. Reposition 

the fuse holder and turn clockwise until fuse 

holder is secure, be careful not to over-tighten. 

Return the switch to the "On" position and

begin operation.

•  To clean, remove the rotors and clean with 

isopropyl alcohol. The housing and lid can be 

cleaned with a damp cloth and a mild 

non-corrosive detergent.

•  After cleaning, ensure all parts are dried thoroughly 

before attempting to operate the unit.

Troubleshooting

Error

Cause

Solution

Unit does not operate

Loose cord. 

Check cord. 

On/Off switch. 

Turn switch to "On" position. 

Lid not closed. 

Close lid until it clicks and locks in 

place. 

Blown fuse. 

Replace fuse. 

Excessive noise or 

vibration

Rotor not balanced. 

Check tube fl uid levels and symmetry. 

Rotor not installed 

properly. 

Remove & reinstall rotor per above 

instructions. 

Foreign object in unit. 

Check centrifuge bowl for loose tubes 

or objects. 

Lid will not open when 

button is pressed

Power is off. 

See above. 

Open lid manually.

Unit stops running 

without pressing front 

button

Loose cord.

Check cord. 

Lid not completely closed.  Close lid until both pins are locked. 

60-minute time limit 

expired. 

Centrifugation cycle is complete. 

1065Y17M_Stuart_Microcentrifuge_Manual_ENGLISH.indd   6

01/11/2019   3:26:53 PM

Содержание Microcentrifuge 8500

Страница 1: ...Microcentrifuge 8500 Operator s Manual RoHS 1065Y17M_Stuart_Microcentrifuge_Manual_ENGLISH indd 1 01 11 2019 3 26 52 PM...

Страница 2: ...t No SCF4 V05 Power Cord 1 US 1 EU 1 UK Operator s Manual 1 each Catalog Numbers 17414 70 SCF4 Maximum speed 8500 rpm Maximum G Force RCF 3884 Capacity 8 place rotor for 0 5 2 0 mL tubes 4 place 8 str...

Страница 3: ...se microcentrifuge tubes and tube strips made from plastic designed to withstand centrifuged forces of at least 3884 g s RCF Always only use original rotors power cord tube adapters and any other spar...

Страница 4: ...to the equipment rating plate to ensure that the plug and fusing are suitable for the voltage and wattage stated The wires in the power cable are as follows 230V 120V BROWN HOT LIVE BLACK HOT LIVE BL...

Страница 5: ...When the Centrifuge has completed the spin cycle an automatic lid unlock will occur If the lid does not open press and hold the button on the front of the unit for one full second the Centrifuge will...

Страница 6: ...isopropyl alcohol The housing and lid can be cleaned with a damp cloth and a mild non corrosive detergent After cleaning ensure all parts are dried thoroughly before attempting to operate the unit Tr...

Страница 7: ...xceed price of product This includes liability claims of loss of use lost profits or any exemplary or punitive damages Repair Should your Stuart Microcentrifuge require service or repair please contac...

Страница 8: ...Class A IEC EN 61326 1 2013 IEC EN 61010 1 2010 Safety requirements for electrical equipment for measurement control and laboratory use Part 1 General requirements Particular requirements for laborato...

Отзывы: