Recalibration
Like all measurement instruments the Pro 360 must periodically be monitored for accuracy.
The Pro 360 should be tested daily with the simple procedure outlined below. If it is found to
be out of calibration, it is easily recalibrated on site using Superset
®
- an eight-step procedure
that realigns the unit through its entire 360° range and takes just a couple of minutes to
perform. And Superset
®
doesn't require any special fixtures, tools, or expertise.
Accuracy Test
Perform this simple test each day before using the Pro 360. Also perform it anytime the
Pro 360 has been dropped or is being used in an environment that varies more than 5°
C (9° F) from the environment in which it was last calibrated.
If your Pro 360 fails this
test, you must recalibrate it using Superset
®
before using the unit further to measure angles.
• Position the Pro 360 with the display
facing you on a clean, flat horizontal
surface. It doesn't have to be exactly
level. Wait 10 seconds so the unit is
completely settled and note the angle
on the display.