6.4 Cleaning and Disinfecting the Sterilizable Light Handle Cover
Cautions
•
The button is not sterilizable.
•
Power off system before cleaning. Make sure that no cleaning fluid enters the system.
•
To avoid damage to plastic parts, do not use any scouring, alkaline, acidic, or alcohol-based cleaning
agents.
•
Allow the light to cool before disinfecting.
•
Handles are shipped non-sterile. Clean and sterilize handles prior to use.
Note: Cleaning and disinfecting must be carried out by trained personnel.
Wipe the surfaces with a cloth and, if required, a mild cleaning solution.
6.4.1 Cleaning
1. Rinse the handle under running lukewarm water, flushing water through all passages.
2. Submerge the entire handle in an enzyme-action detergent prepared according to the manufacturer’s
recommendations. Remove debris and bioburden from all surfaces using a soft brush.
3. Rinse the handle thoroughly under running lukewarm water, flushing water through all passages.
4. Dry the handle with a lint-free cloth and inspect the handle to make sure it is free of debris.
6.4.2 Sterilization
The sterilization methods in the following table have been validated:
Sterilization Method
Temperature
Exposure Time
Gravity-displacement steam sterilization (single-wrapped)
132 °C (270 °F)
10 minutes
Gravity-displacement steam sterilization (double-wrapped)
121 °C (250 °F)
25 minutes
Pre-vacuum cycle (double-wrapped)
134 °C (273 °F)
2 minutes
Note: These cleaning and sterilization instructions are only recommended instructions and
may vary among hospitals depending upon the type of cleaning/sterilization equipment and
possible pathogens in the particular area. These instructions represent exposure time and not
total cycle time. The recommended cleaning/sterilization parameters may be modified based
on standard hospital cleaning/sterilization protocol.