20
RF emissions CISPR 11
Class B
The VisionPro 26” LED Display is suitable for use in all establishments
other than domestic establishments and those directly connected to
the public low-voltage power supply network that supplies buildings
used for domestic purposes, provided the following warning is heeded:
Warning: This system is intended for use by health care professionals
only. This system may cause radio interference or may disrupt the
operation of nearby equipment. It may be necessary to take mitigation
measures, such as reorienting or relocating the system or shielding the
location.
Harmonic emissions IEC61000-3-2
Class D
Voltage Fluctuations/ fl icker
emissions IEC61000-3-3
Complies
Guidance and Manufacturer’s Declaration — Electromagnetic Immunity
The VisionPro 26” LED Display is intended for use in the electromagnetic environment specifi ed below. The customer or the
user of the display should ensure that it is used in such an environment.
Immunity Test
IEC 60601 Test Level
Compliance Level
Electromagnetic Environment--
Guidance
Electrostatic
Discharge (ESD)
IEC61000-4-2
± 6kV contact
± 8kV air
± 6kV contact
± 8kV air
Floors should be wood, concrete,
or ceramic tile. If fl oors are covered
with synthetic material, the relative
humidity should be at least 30%.
Electrical fast
transient/burst
IEC61000-4-4
± 2kV for power supply lines
± 1kV for input/output lines
± 2kV line to ground
± 1kV line to line
Mains power quality should be that
of a typical commercial or hospital
environment.
Surge
IEC61000-4-5
± 1kV diff erential mode
± 2kV common mode
± 1kV diff erential mode
± 2kV common mode
Mains power quality should be that
of a typical commercial or hospital
environment
Voltage dips, short
interruptions and
voltage variations
on power supply
input lines
IEC61000-4-11
• <5%
U
T
(>95% dip in Ut)
for 0.5 cycle
• 40%
U
T
(60% dip in Ut) for
5 cycles
• 70%
U
T
(30% dip in Ut) for
25 cycles
• <5%
U
T
(>95% dip in Ut)
for 5 sec.
• <5%
U
T
(>95% dip in Ut)
for 0.5 cycle
• 40%
U
T
(60% dip in Ut) for
5 cycles
• 70%
U
T
(30% dip in Ut) for
25 cycles
• <5%
U
T
(>95% dip in Ut)
for 5 sec
Mains power quality should be
that of a typical commercial or
hospital environment. If the user of
the transmitter requires continued
operation during power mains
interruptions, it is recommended
that the Wireless Transmitter be
powered from an uninterruptible
power supply or a battery.
Power frequency
(50/60Hz) magnetic
fi eld
IEC 61000-4-8
3.0 A/m
3.0 A/m
Power-frequency magnetic
fi elds should be at levels
characteristic of a typical location
in a typical commercial or hospital
environment.
Note:
U
T
is the AC mains voltage prior to application of the test level.