20
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structures, objects, and people.
(a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless)
telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast,
cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to
fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field
strength in the location in which theVISIONELECT (model 240-030-930) monitor system is used
exceeds the applicable RF compliance level above, the VISIONELECT (model 240-030-930) monitor
system should be observed to verify normal operation. If abnormal performance is observed,
additional measures may be necessary, such as reorienting or relocating the VISIONELECT monitor
VISIONELECT (model 240-030-930) monitor unit.
(b) Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Guidance and Manufacturer's Declaration: Electromagnetic Immunity
VISIONELECT (model 240-030-930) monitor is intended for use in the electromagnetic environment
specified below. The customer or the user of VISIONELECT (model 240-030-930) monitor should
ensure that it is used in such an environment.