Figure 16
Figure 17
Distal Femoral Resection
>
The distal resection level may be altered by
repositioning the Distal Resection Guide in the 2
holes . This will remove an additional 2mm of bone .
>
Once the final resection level is determined, the
distal femoral resection is made . An optional
Modular Capture can be attached to the Distal
Resection Guide .
>
The Triathlon MIS Knee System Instruments are
designed to provide control of the saw blade during
bone resections . When using captures or cutting
through slots, a .050” (1 .25mm) thick blade is used .
14
>
Remove the modular capture and check the
resection for flatness .
>
Remove the Headless Pins in the Distal Resection
Guide using the Pin Puller .
>
Remove the Distal Resection Guide and check the
resection for flatness .
Note: If the “X” pin hole is used, the pin must be
removed prior to repositioning or removing the Distal
Resection Guide .
Femoral
Preparation
Triathlon
®
Knee System
Single-Use Instruments Surgical Protocol