IMPORTANT INFORMATION:  

File in your maintenance records 

  ENGLISH

For use with Total Performance System 
handpieces.

WARNING: 

Before using any tPS component, 

read and understand its instructions and review the 
important User/Patient Safety information provided in 
the tPS Console instruction booklet. Familiarization 
with the total Performance System prior to use is 
important. If you have any questions, contact your 
Stryker Instruments representative or Stryker Customer 
Service at 1-800-253-3210.

WARNING: 

Prior to use, inspect for damage. 

Do not use if damage is apparent.

WARNING: 

Place handswitch in SAFE position 

while changing cutting accessory.

Important Information

Refer to your 

TPS Cleaning, Maintenance, and 

Sterilization Recommendations

 for instrument care 

information.

Installation

NOTE: 

Cord must be removed from  

handpiece to install handswitch.

1. 

Slide safety switch to SAFE position. Figure 1.

2. 

Align white dots and press handswitch band over end of 
handpiece. Figure 1. 

 

NOTE: 

Handswitch band must be flush with handpiece so cord 

can seat properly. Figure 2.

3. 

Plug cord into handpiece. Figure 3.

4. 

Adjust finger rest as needed. Figure 3. the handpiece can be 
held like a pencil. Figure 4.

 

NOTE:

 When using the handswitch with the tPS Oscillating saw, 

the finger rest must be either fully extended or retracted.

Instructions

1. 

Slide safety switch to RUN position.

2. 

Depress lever to regulate variable handpiece speed. 

 

NOTE:

 Maximum speed setting is selected on console.

WARNING: 

Always put the safety switch into the safety 

position to prevent inadvertent running of the handpiece.

To remove:

1. 

Place safety switch in SAFE position.

2. 

Remove cord.

3. 

Grasp band portion of handswitch and slide off back of 
handpiece taking care not to pull back on lever.

RUN 

= run

LOAD 

= load

Safety Switch (shown in 

SAFE position)

White  

alignment dots

Figure 1

Surfaces must  
be flush.

Figure 2

Figure 3

Figure 4

SAFE

0

RUN

I

SAFE

0

RUN

I

Troubleshooting

Problem

Cause

Solution

Handswitch does not work.

Not properly seated on handpiece.

Safety switch in SAFE position.

Reseat.

Place safety switch in RUN position.

Motor does not reach full  
speed.

User may be sideloading switch.

Depress lever centered against 
handpiece.