Now attach the Recon Paddle Trocar
to the T-Handle, AO Medium
Coupling (Fig. 21).
Then, advance them together with the
Recon Tissue Protection Sleeve to the
skin through the hole on the Target
Device labeled (A). Make a small skin
incision and push the assembly
through until it is in contact with the
lateral cortex. Then turn the Knob
clockwise to the position labeled (A)
(Fig. 22).
Remove the Trocar and then insert the
Recon K-Wire Sleeve through the
Tissue Protection Sleeve. Place a Recon
K-Wire into the K-Wire Inserter and
attach it to the T-Handle.
The K-Wire is then man u ally
advanced through the K-Wire Sleeve
until it reaches the subchon dral bone
of the femoral head (Fig. 23).
Alternatively, the K-Wire Inserter can
be attached to a Power Tool and the
Recon K-Wire is inserted to the same
depth.
Fig. 22
Fig. 23
Fig. 21
Operative Technique