Instructions (cont’d)
To Install Battery Pack
To Operate Handpiece
Slide a fully charged battery pack firmly into the
handpiece until the battery latch snaps, indicating the
battery pack is secure (see figure 3).
1.
Figure 3 Install Battery Pack
Test the operation of the handpiece by sliding the
fAST/STD/Safe control to the fAST or STD positions
and squeezing the trigger.
Slide fAST/STD/Safe control to the safe position
until you are ready to use the handpiece.
2.
3.
NOTES:
• This handpiece is designed to accept the Stryker
Large Battery Pack Ref 6215 only. This battery
pack can be charged in the Stryker System 6 Battery
Charger Ref 6110-120 configured with the appropri-
ate battery charger module.
• See the instructions supplied with the battery charger
and/or battery pack for charging details and specifi-
cations.
WARNING:
ALWAYS place the fAST/
STD/Safe control in the safe position while
the handpiece is idle, before installing or
removing an accessory, or when passing
the handpiece to another person. failure to
comply may result in patient and/or health
care staff injury.
CAUTIONS:
• When operating the handpiece, let the blade do the
cutting. Applying too much pressure will bend the
blade and reduce the cutting quality.
• DO NOT stall the handpiece. failure to comply may
damage the electric motor and/or battery pack. If
the handpiece jams, release the trigger immediately.
Remove any obstructions before continuing the pro-
cedure.
• If any power loss is experienced while using a hand-
piece, ALWAYS replace the battery pack with a fully
charged battery pack. failure to comply may result in
a drained or damaged battery pack with a shortened
life.
Slide the fAST/STD/Safe control to the fAST or STD
position to allow the handpiece to operate.
Squeeze the pressure sensitive trigger for variable
speed operation.
Slide the fAST/STD/Safe control to the safe position
when you are finished operating the handpiece.
1.
2.
3.