Service Information
7
HYDRAULIC SYSTEM TROUBLESHOOTING
PROBLEM/SYMPTOM
SOLUTION
Jack will not raise to full height.
Add hydraulic fluid (see below). Check for leaks.
Jack will not hold in raised position.
Close the needle valve completely. If the jack holds,
replace valve (
. If the jack does not hold, re-
place jack.
Jack will not pump up.
Remove excess air (vacuum) in system (see below).
If the jack still will not pump up, replace jack.
Contact Stryker technical support at 1
−
800
−
327
−
0770 for further assistance.
REMOVING EXCESS AIR (VACUUM) FROM THE HYDRAULIC SYSTEM
1. Pump the jack completely up and continue to pump the pedal after the jack reaches the top. This will
force the air through the system and the jack should now pump up.
JACK DESCENT RATE ADJUSTMENT
1. Pump the jack up to full height.
2. The descent rate butterfly valve is located on the manifold assembly. Reach under the base hood and
turn the valve clockwise to decrease the rate of descent. Turn it counterclockwise to increase the rate
of descent.