SPY Portable Handheld Imaging System
Operator’s Manual
21
Figure 4: Video Processor / Illuminator rear panel showing connectors and ports
1.
Main power switch
6
RJ45 port (Communication port)
2.
Device identification label
7
Ext. sync
3.
Potential equalization conductor
8.
SDI 1 video out
4.
Video format switch
9.
SDI 2 video out
5.
RS232 port (Communication port)
10.
DVI video out
CAUTION:
The communication ports on the rear panel are intended for use by NOVADAQ
service representatives only.
Note:
Normal use of the system does not require the user to connect to the potential
equalization conductor. The potential equalization conductor can be used to
equalize across multiple devices or to provide a redundant ground. The potential
equalization conductor is only to be used by qualified personnel. (IEC 60601-1 Ed.
3.2 en:2020 Medical electrical equipment
– Part 1: General requirements for basic
safety and essential performance CONSOLIDATED EDITION, § 8.6.7)
A medical-grade color video monitor can be connected via the SDI and DVI
connectors on the rear panel of the VPI (Figure 4).
Appendix C
for specifications of the video output signals to
ensure compatibility of the selected video monitor. Consult the operator instructions
provided by the manufacturer of the video monitor for details on operation and
adjustment of the video monitor.
Содержание SPY PHI HH9000
Страница 1: ......
Страница 6: ...iv 4 0003045 Rev G This page intentionally left blank...
Страница 8: ...2 4 0003045 Rev G This page intentionally left blank...
Страница 36: ...30 4 0003045 Rev G This page intentionally left blank...
Страница 58: ...52 4 0003045 Rev G This page intentionally left blank...
Страница 102: ...96 4 0003045 Rev G This page intentionally left blank...
Страница 106: ...100 4 0003045 Rev G This page intentionally left blank...
Страница 110: ...104 4 0003045 Rev G This page intentionally left blank...
Страница 111: ......
Страница 112: ......