Operation

Checking pressure setting by hand

Check the pressure setting by hand to make sure that the patient receives the proper therapy. 
Perform a hand check with the patient in the worst case position, which is the side-lying position. 
This position maximizes the pressure at the sacrum.

Figure 1: Under inflation

Figure 2: Proper inflation

1. 

Slide your hand, palm up with fingers flat, between the product and the mattress at the 
patient’s sacrum. 

Notes: 

• 

Do not lean on the product or lift at the side as these actions can lead to false readings.

• 

Adjust the pressure setpoint until you can no longer feel the patient’s sacrum.

2. 

If you can feel the bony prominence (Figure 1), inflate the product.

3. 

Increase the pressure up to 3mmHg by pressing the firmness button on the pump.

4. 

Wait two minutes and repeat the hand check until you have adequate inflation  
(Figure 2).

Note: 

Check the patient at regular intervals as directed by hospital protocol.

Introduction

This manual assists you with the operation of your Stryker product. Read this manual before 
operating this product. Set methods and procedures to educate and train your staff on the safe 
operation or maintenance of this product.

Notes

:

• 

This manual is a permanent part of the product and should remain with the product even if 
the product is sold.

• 

Stryker continually seeks advancements in product design and quality. This manual contains 
the most current product information available at the time of printing. There may be minor 
discrepancies between your product and this manual. If you have any questions, contact 
Stryker Customer Service or Technical Support at 1-800-327-0770.

Product description

The Stryker Model 2790-100-000 

SPR Plus

®

 

Overlay and Model 2863 

AIR+

™ pump together 

create a low air loss system to manage microclimate of heat and humidity that helps keep 
moisture away from the patient’s skin. Major components of the system include the pump, hose, 
hose connector, and overlay. The pump provides air to the overlay that contains micro vents to 
dispel small volumes of air.

Intended use

This product is intended to assist in the prevention and treatment of pressure injuries stages 
(1, 2, 3, 4, unstageable and deep tissue injury or all pressure injuries) as well as helps manage 
microclimate of the patient’s skin.  It is recommended to be implemented in combination with 
clinical evaluation of risk factors and skin assessments made by a healthcare professional. The 
product is for use by trained health care professionals. This product is for use in acute (critical 
care, step down, med/surg, sub-acute, post anesthesia care unit (PACU)), or long term care 
settings. This product is for single patient use.

This product is not intended to be used: 
• 

When spinal stability is a concern

• 

To support a patient in a prone position

Contraindications

None known.

Expected life

This product has a 30 day, single patient use, expected life under normal use conditions.

Introduction

Have the lot batch code (E) available when you call Customer Service. Include the lot batch 
code in all written communication.

Product illustration

A

B

C

D

E

A

Head end

C

Corner straps

E

Lot batch code

B

Hose connection

D

Foot end

Cleaning

1. 

Clean the product with a clean damp cloth, soap, and water.

2. 

Wipe dry.

3. 

Dispose of the overlay according to your local waste management policy.

Notes: 

• 

Contact with disinfectant cleaners will cause degradation to the 

SPR Plus 

product. Avoid 

disinfectants, such as solvents, alcohol based products, or Virex TB germicidal products.

• 

Iodine based disinfectants will stain the product. Stains do not affect the patient or the use 
of the product.

Setup the overlay

1. 

Remove the product from the plastic bag. Place the product across the bed with the hose 
connection on the right side of a supine patient.

2. 

Stand at the foot end of the bed, unroll the cushion until you see the air valve on the right 
side.

3. 

Place the corner straps around the mattress cover.

4. 

Make sure that the hose connection is easy to access. The air holes should face up for 
airflow.

5. 

Place a sheet and incontinence pad, if required, over the cushion. See 

Adding or changing 

linens

.

6. 

Inflate the Overlay with or without the patient on the product.

Setup the pump and hose

1. 

Bring the product to room temperature. 

2. 

Place the 

AIR+

 pump on the foot panel of the bed or secure location. 

3. 

Plug the power cord into a hospital grade outlet.

4. 

Screw the pump hose connector cap to the product connector.  Do not cross the threads.

5. 

Select the 

SPR Plus 

mode on the control panel of the 

AIR+

 pump.

6. 

Press the mmHg button on the control panel. Adjust the pressure setting as required.

Note: 

See the 

AIR+

 pump Operations Manual for more information.

Adding or changing linens

• 

Apply the linens loose to avoid a hammock effect.

• 

Use minimal layers of sheeting and incontinence pads.

• 

Change the linens per your hospital procedures.

Contact information

Contact Stryker Customer Service at 1-800-327-0770.

Stryker Medical
3800 E. Centre Avenue
Portage, MI 49002  
USA

To view your operations or maintenance manual online, see https://techweb.stryker.com/.

Symbols

General warning

Caution

Consult instructions for use

Model number

Batch code

No sharp objects

Safe working load

1

Single patient use

Foot end

Manufacturer

Plastic #3- Polyvinyl Chloride (PVC or V)

Specifications

Safe working load

350 lb

 158.8 kg

Material

Vinyl

Length

78 in.

198 cm

Width

34 in.

86 cm

Weight

6 lb

2.72 kg

Classified

Class II, not classified for protection against harmful 
ingress of liquid

Compatible pump

 Model 2863 

AIR+

 pump

Environmental conditions

Operation

Storage

Operating ambient temperature 
range and humidity limits

50   F

(10  C)

 0

 0

  104   F

(40  C)

 0

 0

30%

75%

Storage temperature range and 
humidity limits

140 °F
(60 °C)

-40 °F

(-40 °C)

10%

95%

Stryker reserves the right to change specifications without notice.