Operation
Footboard control panel - Scale (Continued)
To weigh a patient:
1.
Press and hold the
Scale
button (A) (see
Footboard control panel - Scale on page 44
) until
Release Button
appears (
).
Figure 28: Release button
2.
Release the
Scale
button.
Note:
Do not touch the product when you weigh the patient (
Figure 29: Do not touch bed
A confirmation notification indicates that weighing the patient was successful (
Figure 30: Patient weight
Note:
To clear the patient weight from the display, press the
Scale
button again. The patient weight is still recorded in
the weight log.
Notes
•
Always calibrate the bed after adding a support surface or mattress to the bed frame.
•
Always set the scale to zero before putting a patient on the product.
Setting the scale to zero
The zero function clears all of the stored values from the weight log, change patient weight, and gain or loss.
Note:
Always set the scale to zero before putting a patient on the product.