Stryker Reprocessed Ultrasonic Focus Скачать руководство пользователя страница 3

Reprocessed Ultrasonic Focus™ Curved Shears 

 

   Avoid contact with any and all metal or plastic instruments or objects during instrument activation. Contact with staples, 

clips, or other instruments during instrument activation may result in premature blade failure, resulting in generator solid 
tone or instrument error. 

  Avoid incidental and prolonged activation against solid surfaces (such as bone), which may result in blade heating 

and/or blade failure. 

  Take care to avoid application of pressure between the blade and tissue pad without tissue in between them as this can 

result in damage to the instrument.  Both conditions may cause a system failure signaled by a continuous beep when 
either of the foot pedals or hand controls buttons are depressed.  

 

 

When the instrument blade is activated the blade tip is active and will cut and coagulate tissue.  Care must be taken to 

avoid unintentional contact between the blade surfaces and surrounding tissues.

 

  The instruments Minimum starting power level defaults to power level 3. 

  Verify hemostasis after withdrawing instrument. If bleeding is still observed, employ appropriate techniques to achieve 

hemostasis. 

  When  using the Reprocessed  Ultrasonic Curved  Shears  on solid organs,  use caution. Due to the limited ability of the 

shears  to grasp large portions of solid organs and occlude vascular structures of this nature, hemostasis may not be 
predictable and may require additional measures for coagulation.  

  Instruments or devices which come into contact with bodily fluids may require special disposal handling to prevent 

biological contamination.  Special disposal handling techniques may be needed to prevent biological contamination if 
instruments or devices come into contact with bodily fluids. 

 
Directions for Use 

The package label is detachable and may be affixed to the medical record of the patient. 

 

 
Assembly 

1. 

Inspect the instrument and package before opening. The contents of the package are sterile if the packaging has 
not been compromised. If the package is damaged or if it was opened and the instrument was not used, return 
the instrument and package to Stryker Sustainability Solutions. 

2. 

Remove the instrument from the package and place it in a sterile work area using aseptic technique. 

3. 

Inspect the instrument for overall condition and physical integrity. Do not use the instrument if any damage is 
noted. Return the instrument and packaging to Stryker Sustainability Solutions if it is not in acceptable condition 
for surgery. 

4. 

Attach the Blue Grip Assist on the Hand Piece to the instrument by rotating the instrument onto the Hand Piece in 
a clock wise rotation as viewed from the  distal end of the instrument (finger tight only).  

5. 

The Blue Torque wrench is used to tighten the instrument onto the Hand Piece.  Use the Grip Assist to hold the 
Hand Piece and not the instrument handle when applying the Blue Torque Wrench. Turn the Blue Torque Wrench 
clockwise while holding the Hand Piece with the Grip Assist until it snaps twice indicative that sufficient torque has 
been applied to secure the instrument.  

Note: 

Do not use any other means than the Torque Wrench and Grip Assist to attach or detach the device from 

the Hand Piece. 

Note:  

Do not hand torque, damage to the Hand Piece may occur. 

6. 

Carefully remove the Blue Torque Wrench and Blue Grip Assist from the Hand Piece.  Do not dispose of the Blue 
Torque Wrench and Grip Assist until completion of the procedure as it is used for removal of the instrument from 
the Hand Piece following the procedure. 

7.  Note: 

Take care to avoid damage to the blade while sliding the Blue Torque Wrench onto or off the shaft.  

8.  Note: 

Take care to avoid injury from the blade tip while sliding the Blue Torque Wrench onto or off of the shaft.  

 
Operation 

Refer to the Harmonic™ Generator 300 Operator’s Manual (GEN04) for Hand Piece attachment and system 
operations instructions. 

1. 

Before beginning the procedure, verify compatibility of all instruments and accessories. 

2. 

The generator power may be turned ON after the assembled Hand Piece and Instrument have been connected to 
the generator. 

a. 

Note: 

Ensure the Hand Piece and instrument are connected to the generator before turning on the power. 

3. 

Select the desired MIN power level using the INCREASE and DECREASE buttons on the generator.  

a. 

Note: 

MIN power is defaulted at power level 3.   

4. 

Higher generator power level is warranted for greater tissue cutting speed and for greater coagulation use a lower 
generator power level.  The amount of energy delivered to the tissue and resultant tissue effects are a function of 
many factors, including the selected power level,  blade characteristics, grip force, tissue tension, tissue type, 
pathology, and surgical technique.  

Содержание Reprocessed Ultrasonic Focus

Страница 1: ...ral U S A law restricts this device to sale by or on the order of a physician STERILE Explanation of Icons Sterilized by Ethylene Oxide Gas Date of Reprocessing Use by Date Product Code Do Not Reuse See Instructions For Use Stryker Sustainability Solutions Inc 2006 2011 Phoenix AZ 85044 sustainability stryker com 888 888 3433 ...

Страница 2: ... hazards consult the medical literature Prior to the initiation of the procedure verify compatibility of minimally invasive instruments and accessories when they will be used together since there may be variations from manufacturer to manufacturer The use of these instruments requires a thorough understanding of the techniques and principles of laser electrosurgical and ultrasonic procedures Inapp...

Страница 3: ...ect the instrument for overall condition and physical integrity Do not use the instrument if any damage is noted Return the instrument and packaging to Stryker Sustainability Solutions if it is not in acceptable condition for surgery 4 Attach the Blue Grip Assist on the Hand Piece to the instrument by rotating the instrument onto the Hand Piece in a clock wise rotation as viewed from the distal en...

Страница 4: ... power level is activated by pressing the right foot pedal of the footswitch or distal hand control button MAX 11 Note Premature blade failure could occur from scratches on the blade a When instrument is in use avoid accidental contact with other instruments b Use only the Torque Wrench to attach or detach the instrument from the hand piece 12 Close the clamp arm by closing the finger rings togeth...

Страница 5: ...al numbers or other identification marks have been removed or destroyed If Stryker determines in its reasonable discretion that the claimed defect or non conformance in the product is excluded from warranty coverage as described hereunder it will notify the customer of such determination TO THE FULLEST EXTENT PERMITTED BY LAW THE EXPRESS WARRANTY SET FORTH HEREIN IS THE ONLY WARRANTY APPLICABLE TO...

Отзывы: