INTERVAL

INSPECTION CRITERIA

ACTION

Before
Use

Inspect equipment for damage, wear, or 
missing components.

Inspect power cord and twist cable for 
cuts or punctures.

Inspect headlight lens for scratches.

Inspect the power pack housing for cracks.

If damage is
apparent, replace
the equipment.

Lorem ipsum

Instructions

Inspection and Testing

Only individuals trained and experienced in the maintenance of reusable 
medical devices should inspect and test this equipment.

DO NOT use any equipment if damage is apparent or the inspection criteria 
are not met.

DO NOT disassemble or service this equipment, unless otherwise specified.

Perform recommended inspection and testing as indicated in the Inspection 
and Testing section.

ALWAYS operate the equipment within the specified environmental condition 
values. See the Specifications section.

This equipment is suitable to use in a professional healthcare facility environment.

DO NOT cut or puncture the twist cable.

ALWAYS turn the light source power off before connecting or disconnecting 
the twist cable to or from the LED surgical headlight.

ALWAYS hold the strain relief near the end of the twist cable when connecting 
or disconnecting the twist cable from the light source.

DO NOT scratch the LED surgical headlight lens.

DO NOT stretch or kink the twist cable. Failure to comply will damage the 
twist cable and reduce the light output permanently.

NOTE: 

Headband position will vary from person to person depending on head 

size and personal comfort preferences.

CAUTIONS:

To Don a LED Surgical Headlight

Connect the LED surgical headlight power cable to the appropriate twist cable 
(figure 2). See the Accessories section.

NOTE:

 To disconnect the power cable, twist the connector sleeve back before 

separating the connector plug and receptacle (figure 2).

DO NOT connect the twist cable to the power pack before donning the LED 
surgical headlight.

1.

Loosen the headband adjustment knob on the top and back of the LED 
surgical headlight (figure 3).

2.

Put the LED surgical headlight on your head and turn the headband knob on 
the top and back to secure the LED surgical headlight (figure 4). When using 
loupes, put them on after this step.

NOTE:

 Make sure the LED surgical headlight fits securely on top of your head 

and that the light is coaxial to your line of sight (figure 5).

3.

To Use a LED Surgical Headlight

Plug the twist cable connector into the jack on the power pack (figure 6).

Press the power button on top of the power pack (figure 7).

Press the variable control buttons (”+” and “-”) to adjust the light intensity. 
See the 

Troubleshooting

 section if required (figure 8).

1.

2.

3.

To Doff a LED Surgical Headlight

Turn the power pack off.

Loosen the headband knobs on the top and back of the LED surgical headlight 
and remove the LED surgical headlight from your head.

Clean the LED surgical headlight and cord. See the 

Cleaning

 section. To clean 

the PP4 Power Pack, see the instructions for use supplied with the PP4 
Charging Base Station.

1.

2.

3.

NOTE:

 For service, contact your 

Enova's

 sales representative or call 

Enova's

 customer service.

NOTE:

 If any component must be discarded, see the 

Disposal / Recycle

 

section.

WARNINGS:

WARNINGS:

Содержание Quantum LED Surgical Headlights

Страница 1: ...Quantum LED Surgical Headlights Instructions For Use 2021 07 Quantum Cool LED Surgical Headlight Quantum Warm LED Surgical Headlight ENGLISH EN ...

Страница 2: ...T look directly into the LED surgical headlight when illuminated DO NOT shine the light directly into the patient eyes or face Due to the intensity of light eye injury may result Patient s eyes should be appropriately protected in surgeries involving the face or head After a few minutes of operation the headlamp housing may become warm to the touch This is normal and functions to dissipate heat fr...

Страница 3: ...his bracket is used to adjust the light beam to fit coaxially to the line of sight Headband Adjustment Knobs Customize the fit by turning the knob clockwise to tighten and counterclockwise to loosen the headband Quantum Twist Cable This cable routes light from the power pack to the LED surgical headlight Comfort Pads These removable pads providea comfortable fit C A B D E ...

Страница 4: ...ng on head size and personal comfort preferences CAUTIONS To Don a LED Surgical Headlight Connect the LED surgical headlight power cable to the appropriate twist cable figure 2 See the Accessories section NOTE To disconnect the power cable twist the connector sleeve back before separating the connector plug and receptacle figure 2 DO NOT connect the twist cable to the power pack before donning the...

Страница 5: ...www goenova com 5 F20006 Rev E EN NOTE It is recommended the operator wears gloves while performing these steps Instructions 1 5 6 3 7 4 8 2 ...

Страница 6: ...values throughout its useful life See the Specifications section To ensure the longevity performance and safety of this equipment use of the original packaging material is recommended when storing or transporting this equipment Disposal Recycle WARNINGS ALWAYS follow the current local recommendations and or regulations governing environmental protection and risks associated with recycling or dispo...

Страница 7: ...tmospheric Pressure Limitation Operation 10o C 30o C 30 75 70 kPa 106 kPa Storage and Transportation 5o C 50o C 10 75 50 kPa 106 kPa Symbol Definition Definition Indication Color General Warning Sign Yellow Refer to Instruction Manual Booklet Blue Catalog number Quantity Lot number Storage Temperature Limitation Transportation Temperature Limitation Operation Temperature Limitation Storage Transpo...

Страница 8: ...2021 07 www goenova com F20006 Rev E ...

Отзывы: