6. Using a 3/16” hex wrench, remove the two flat head cap screws (A) that secure the upper IV pole clamps (C) to the chair
frame. Remove the clevis pins (D) and clamps (C). Save the screws, clevis pins, and clamps. Discard the IV pole (E).
N
No
otte
e -- Hold the IV pole assembly while removing the flat head cap screws to keep the IV pole from falling.
IIV
V p
po
olle
e iin
nsstta
alllla
attiio
on
n
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
3/16” hex wrench
•
Small slotted screwdriver
•
3/8” drive torque wrench (ft-lb)
•
3/8” drive torque wrench (in-lb)
P
Prro
occe
ed
du
urre
e::
N
No
otte
e -- An umbrella nut (0018-045-000) is also required when replacing the IV pole.
1. Remove the seat back (
2. Insert the supplied umbrella nut (0018-045-000) (J) into the top of the IV pole (E) (
(page 22)). The
top of the umbrella nut should be about 1/2” from the top of the IV pole.
N
No
otte
e -- The umbrella nut may be installed before shipment. Make sure that the IV pole has the umbrella nut in place
before installing the IV pole topper.
3. Insert the supplied tube plug (0037-225-000) (H) into the bottom of the IV pole (E).
4. Position the IV pole topper (F) on the IV pole (E) and use the palm of your hand to tap the topper until the topper is
seated on the IV pole (E).
5. Using a 3/16” hex wrench, install the supplied socket head cap screw (0004-672-000 (G) to secure the IV pole topper to
the pole. Using a 3/8” drive torque wrench (in-lb), torque the screw to 25 ± 5 in-lb.
6. Using a 3/16” hex wrench, remove the two flat head cap screws (A) that secure the upper IV pole clamps (C) to the chair
frame. Remove the clevis pins (D) and clamps (C). Save the screws, clevis pins, and clamps. Discard the IV pole (E).
7. Insert the supplied clevis pins (0026-349-000) (D) into the front of the supplied upper IV clamp (removed in step 6).
8. Install the front of the upper IV clamp and clevis pins on the chair frame.
9. Install the supplied IV pole into the front of the upper IV clamp and clevis pin.
N
No
otte
e -- Hold the IV pole in position while installing the back of the upper IV clamp to keep the pole in place.
10. Using a 3/16” hex wrench and the supplied flat head cap screws (0001-210-000) (A), install the back of the upper IV
clamp onto the IV pole and front of the upper IV clamp. Using a 3/8” drive torque wrench (ft-lb), torque the screws to 13
± 3 ft-lb.
11. Using a 3/16” hex wrench and the supplied flat head cap screws (0001-210-000) (A), install the supplied lower IV clamp
front and back onto the IV pole. Using a 3/8” drive torque wrench (ft-lb), torque the screws to 13 ± 3 ft-lb.
N
No
otte
e -- Align the lower IV clamp, so the clamp is flush with the bottom of the IV pole.
12. Reinstall the seat back (
13. Verify proper operation before you return the product to service.
IIV
V p
po
olle
e tto
op
pp
pe
err rre
ep
plla
acce
em
me
en
ntt
T
To
oo
ollss rre
eq
qu
uiirre
ed
d::
•
3/16” hex wrench
•
3/8” drive torque wrench (in-lb)
1460-009-002 Rev B.0
23
EN